8.1.15 No load voltage (according to IEC/ EN 60974-4)
No load voltage Up (a.c. peak or d.c. peak)
Range
Resolution
Accuracy
Up
0.0 V
… 199.9 V
0.1 V
(5 % of r 5 D)
No load voltage readout Urms (a.c. r.m.s.)
Range
Resolution
Accuracy
Urms
0.0 V
… 139.9 V
0.1 V
(5 % of r 5 D)
Measurement circuit according to EN 60974-4
– Figure 4.
W-output* .............................. a.c., d.c.
Rated no-load voltage* ......... Custom, -
, 20 V … 99 V
Limit (Up)* ............................. U
0
p
15 %
Limit H (Up)* ......................... a.c. output: Off, 68 V, 113 V, 141 V
d.c. output: Off, 113 V, 141 V
Limit (Urms)* ......................... Off, 48 V, 80 V, 100 V
Test terminals ....................... Test socket (A 1422: 16A-5p, 32A-5p or 16A-3p),
Test terminals W1 / W2
*