A 1322 / A 1422
Measurements with A 1422
26
6.4 Insulation resistance (supply circuit of Class II equipment to
accessible surfaces)
Refer to master instrument Instruction manual, chapter
Single test
measurements
, subchapter
Insulation resistance - Riso (Welding equipment)
for detailed information.
Procedure:
Item
Comment
Prepare test system.
See
chapter 4.3 Working
Connect power supply plug of tested welding
machine into proper test socket.
Connect P/S probe to accessible isolated
conductive part of the welding machine.
Select Riso (Rw) test function on the master
instrument.
Set Test parameter LN (Class II)-P/S.
Start measurement.
Wait until measurement is finished or stop the
measurement.
Check result(s).
Store result(s).
Optional
Figure 6.4:
Insulation resistance (supply circuit of Class II equipment to accessible
surfaces)
Notes:
Consider any warning on the display before starting the measurement!
Do not touch or disconnect the DUT during the insulation resistance
measurement or before it is fully discharged!