24
SySTEM
Figure 26 Configuration tab screens
•
Calibration. Select to start the calibration process (
) for water activity or
moisture content (
). Resetting factory defaults is described in
.
Figure 27 Calibration option
•
Temperature. Select to set the AQUALAB TDL to the desired temperature using UP and
DOWN. When finished, select SAVE (
The AQUALAB TDL may be set between 15 and 50 °C in 1.0 °C intervals. The default
temperature is 25 °C.
Figure 28 Temperature option
•
Temp Eq. Select to set the acceptable temperature difference between the AQUALAB TDL
block and the sample before starting a water activity measurement (
The range is 0.1 to 4.0 °C. A setting of 4.0 °C begins the measurement when the sample is
<4.0 °C above or below the block temperature. A setting of 0.1 °C begins the measurement
when the sample is <0.1 °C above or below the block temperature.