SERVICE
Figure 56 Calibration saved confirmation
16. Select
OK to continue.
17. If the calibration is canceled at any point, a Calibration Canceled confirmation will
appear. Select
OK to continue (
).
Figure 57 Calibration canceled confirmation
18. Measure a 0.25 or 0.50
a
w
verification standard in the Measurement tab.
19. Verify the AQUALAB TDL reads within ±0.005
a
w
at a given temperature.
20. Measure a 0.76
a
w
verification standard.
If both verification readings are within ±0.005
a
w
, the instrument is ready to begin
testing.
If incorrect verification standard readings persist, contact
.
4.1.4 RESTORE FACTORy DEFAULTS
To restore original calibration settings, do the following:
1. Navigate to the Configuration tab using
MENU.
2. Select
Calibration
.
3. Select
Defaults
to access the Restore Factory Defaults routine (
Select
CANCEL to return to the Configuration tab.