48
SERVICE
6. Visually inspect the sample chamber and sensors for contaminants, including moisture.
If necessary, repeat the cleaning process using new KIMWIPES tissues.
7. Let stand for 5 min to ensure the sample chamber is dry.
8. After cleaning the AQUALAB TDL, check the instrument performance and correct for any
linear offset that may have occurred during the cleaning process.
a. Run a sample of the activated charcoal pellets from the AQUALAB Cleaning Kit.
This cleans the air inside the chamber, helping it come back to a stable sampling
environment.
b. Verify the linear offset against known verification standards as described in
c. If a linear offset has occurred, adjust for linear offset.
If the instrument is still not reading samples correctly, contact
.
4.3 MAINTENANCE PACKAGES
METER offers maintenance and calibration packages to ensure AQUALAB is functioning to its
highest standard (
).
Table 5 Maintenace package options
Package
Basic Calibration Service
As-found inspection
Available
Replace old/damaged parts
Separate charge per part
Instrument cleaning
Included
Factory calibration
Included
Extended 1-year warranty
Not included
Loaner instrument
Available
Replacement parts can also be ordered from METER. Contact
.
4.4 REPAIR
METER repairs manufacturer defects and instruments within the 1-year warranty at no
charge. Repairs outside of the warranty window are charged based on cost of parts, labor,
and shipping. An extra fee may be charged for rush work. Contact
for an
estimated repair cost.
METER has loaner instruments available for a fee while the AQUALAB TDL is being serviced.