EN |
Care and maintenance
22
2
Tilt the locking bracket on module 2 upwards, remove it and place it on a fl at, non-slip surface.
3
Open the 4 yellow clip fasteners on module 2, hold the collection container and lift off the upper part.
4
If the pump sieve (1) is dirty, pull it off, clean it over a collection receptacle and put it back on the suction
housing of the pump.
Wipe the probes in module 2 (3) with a paper towel. For easier cleaning of module 2, the air brake (2)
can also be removed.
When assembling, pay attention to the position markings (arrow on the air brake, notch on module 2)!
1
3
2
5
Place the cleaned and properly assembled module 2 on the new collection container.
Pay attention to the FRONT marking on the collection container!
Close and dispose of the full collection container according to instructions (see 7.2.2. Disposal of the
full collection container)
New
Old
6
Close the 4 yellow clip fasteners on module 2.
Clean the sealing rings on the left and right in the supporting element with a damp paper towel and
grease them with multi-silicone grease (or Vaseline).
7
Carefully push module 2 back into the holder and close the locking bracket.