Preparation for use
| EN
17
5.4.1. Electrical connections
6
5
1
4
3
2
1
Main board
2
Diagnostic board
3
Module 2
4
Module 3
5
External display
6
Hose storage switch and suction system isolating relay
SV1
Connector for centrifuge inlet probe
SV2
Connector for centrifuge motor
SV3
Connector for power supply (24 V AC)
SV4
Connector for external display
SV5
Connection to the diagnostic board (power supply for motor and magnetic valve in
module 2)
SV6
Connection to the diagnostic board (probes in module 2)
SV7
Connector for suction signal (12 - 24 V AC or DC)
F1
Main fuse MULTI SYSTEM TYP250/T 3, 15 A/UN 250 V/ ICN 35 A
F2
Fuse MST250/T 1.6 A/UN 250 V/ ICN 35 A for pump motor in module 2
i
Removing the jumpers on J1, J2 and J3 causes a better response of the probes at low water conductivity.
Danger:
The supply voltage must be taken from a safety transformer that complies with the requirements of IEC 601-1 / VDE 0750 Part 1 / DIN
EN 60601-1 and IEC 60742 + A1 / DIN EN 60742.
Danger:
Fuses may only be replaced by the same type!