Maintenance
This section provides information about routine maintenance of the AEROCET-380. The
case of the AEROCET-380 should never be removed or opened for any reason.
Opening
or removing the case of the AEROCET-380
voids the warranty and may result in
exposure to laser radiation, which can cause eye injury.
10.1 Suggested Periodic Maintenance Intervals
The table below shows the service schedule for the AEROCET-380.
Maintenance Item
Suggested Period
Flow Audit/Calibration
Monthly
Replace PUMP filter and PURGE filter
Yearly
Factory service, recalibration, and optical system cleaning
Every 24 months
10.2 Flow Audit and Calibration
This procedure is discussed in Section 5.3.4.
10.3 PUMP and PURGE Filter Changes
The AEROCET-380 PUMP FILTER and PURGE FILTER are the two filter cartridges located
on the side panel of the instrument. They can be removed by unscrewing the black
aluminum filter holders with a coin or flat head screwdriver using the slot in the face of the
holder. The expected lifetime of the two filters is greater than one year, but in heavy
particulate areas they may need to be replaced more often.
The 5 micron PUMP filter keeps any large particles from entering the sample pump. The
0.01 micron PURGE filter cleans the recirculated purge air which is used to keep the optics
clean during normal operation.
Pump Filter
Purge Filter