034E-9800 REV A
Page 16
J.
Refer to Figure 4-2. Feed Pot/Mount Assembly wires into center hold in sensor
stem. Note the location of the flat on the sensor stem as re-alignment of the
Pot/Mount Assembly is critical to re-establish factory calibration.
K.
Seat the unit onto the sensor stem and slowly tighten the set screw as you feel
that the flat alignment has been accomplished.
L.
Spin the Cup Assembly to check for freedom of rotation. There should be a small
amount of up and down play (Approximately .010”).
M.
Re-open sensor body; connect pot wires in terminal block, matching wire colors
as in Step B.
N.
Prior to final closure of sensor body, inspect inside. Ensure that wires are routed
as to prevent pinching or over-straining. Make sure wires do not get between
spacer and interior of sensor body. Close sensor and secure with screw.
O.
Unit is now ready for service.
Flat for set screw