![Met One Instruments 034E Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/met-one-instruments/034e/034e_operation-manual_1775881010.webp)
034E-9800 REV A
Page 10
4.0
MAINTENANCE AND TROUBLESHOOTING
4.1
General Maintenance Schedule
6-12 month intervals:
A.
Inspect the sensor for proper operation per Section 3.0
B.
Replacement of Wind Speed Sensor bearing in extremely adverse environments.
12-24 month intervals:
A.
Replacement of Wind Speed Sensor bearings.
24-36 month intervals:
A.
Recommended complete factory overhaul of sensor.
*Schedule is based on average to adverse environments.
TABLE 4.1
TROUBLESHOOTING TABLE
Symptom
Probable Cause
Remedy
No WS Sensor output
Faulty bearings
Replace bearings (REF. 4.2)
No WD Sensor output
Faulty pot assembly
Replace pot assy. (REF 4.3)
No WS Sensor output
Faulty reed switch
Replace reed switch (REF 4.4)
No WS or WD output
Faulty cable
Check cable and connections
Содержание 034E
Страница 2: ...034E 9800 REV A Page 2 ...
Страница 20: ...034E 9800 REV A Page 20 FIGURE 4 7 EXPLODED PARTS VIEW OF 034E ...