Manual - Appendix

b)

Operation and handling of the 

opto

LAN

Set up the devices at the dedicated places. The transceiver 

connected to the PC can be powered by the power supply (optional) 
all the time, if there is no need for battery supply (EMC issues). The 

transceiver inside the chamber being connected to the DUT does 
not have to be powered by a power supply during testing (might get 

damaged)! It has integrated batteries to run the system.

Connect the transceivers with standard ethernet cables (without 

crossing) to the PC and the DUT. Inside the absorber lined chamber, 
the ethernet cable should be as short as possible with a good 

shielding (e.g. CAT 6 or higher) to avoid rf coupling into the system.

Connect the optical fiber (Out => In; In => Out). Only use 

62,5/125µm multimode fibers.

Connect the electric cables

Turn on all devices (power push button, see Fig. a.1); the order does 
not matter

The transmission starts automatically after a short initialization. 

If the transmission suddenly stops after a long duration of measurement, 

check the 

Info

 LED of the transmitter (see Fig. a.1).

If the battery power falls below 5,2V, the  

Info LED

  is switched on. The 

system should be reloaded soon. Below 4,5V, the system is turned off 
automatically.

The measurements can be extended by using the optional battery pack 
(BP-60)   with   connector   cable   or   a   power   supply   certified   by   mk-

messtechnik. The external supply can be connected to the system any time 
(parallel). The connection to the internal battery is decoupled with a diode.

Only use the battery pack and connector cables from mk-messtechnik! 
Others might lead to a damage of the system!

Check info LED if 
transmission stops 
suddenly!

Only use battery 
packs and connector 
cables provided 
from mk-
messtechnik. Other 
modules influence 
EMS-performance 
and might damage 
the opto-system!

The use of the 
external filter is 
essential for EMS-
tests!

Use short, shielded 
ethernet cables with 
1:1 pinning!
(no crossing!)

Distributed by: Reliant EMC LLC, 3311 Lewis Ave, Signal Hill CA 90755, 408­916­5750, www.reliantemc.com

 

Distributed by: Reliant EMC LLC, 3311 Lewis Ave, Signal Hill CA 90755, 408­916­5750, www.reliantemc.com

 

Содержание optoLAN

Страница 1: ...Manual optoLAN Digital optical transmitter for Ethernet 10 100 Mbit signals Distributed by Reliant EMC LLC 3311 Lewis Ave Signal Hill CA 90755 408 916 5750 www reliantemc com...

Страница 2: ...Distributed by Reliant EMC LLC 3311 Lewis Ave Signal Hill CA 90755 408 916 5750 www reliantemc com...

Страница 3: ...teristics 4 3 Field of application 5 4 Maintenance 5 5 Trouble shooting 7 6 Accessories Options 8 7 Contact 8 Appendix Details and operation A1 Distributed by Reliant EMC LLC 3311 Lewis Ave Signal Hil...

Страница 4: ...gital optical system optoLAN can be used to optically transmit ethernet 10 100 Mbit signals Because of the optical transmission the system is very robust against EMS electromagnetic susceptibility It...

Страница 5: ...automatically Afterwards charge the devices as usual The devices have to be turned off before connecting to the charger If this is disregarded the system might get damaged Fig 4 1 shows the pinning of...

Страница 6: ...a longer time Do not use cleaning agents or solvents to clean the devices only use a slightly moistened soft cloth Do not open the devices as there are no parts inside which have to be maintained The...

Страница 7: ...5 125 m Charge batteries Turn on all devices Transmission stops Low battery No optical signal at receiver System turned off Charge batteries Check for light at optical output Replace optical fiber Tu...

Страница 8: ...LWL 2 xm x length in m duplex External batteries BP 60 6V 4Ah Connector cable for BP 60 AK BP Length approx 15cm Charger with connector plugs CH 5 Standard charger Manual MA optoLAN German or english...

Страница 9: ...tical connector FSMA Optical In Out for duplex multimode fiber 62 5 125 m Fig a 2 shows the rear side of the transceivers with RJ45 connector standard ethernet pinning RJ45 plug LAN I O 10 100Mbit s f...

Страница 10: ...ts automatically after a short initialization If the transmission suddenly stops after a long duration of measurement check the Info LED of the transmitter see Fig a 1 If the battery power falls below...

Отзывы: