opto
LAN
Date:
11/15/11
Manual
Page 5
3
Field of application
•
Transmission of ethernet signals up to 10/100Mbit/s during EMC-
tests
•
Transmission of ethernet signals over long distances without signal
loss (up to 100m or more, depending on timing requirements)
•
Handle ground potential problems
4
Maintenance
Recharge batteries after use with the enclosed charger. To prevent a lazy
battery effect, discharge the devices every 5 times completely by using the
automatic switch off (Leave the system on, until it turns off automatically).
Afterwards, charge the devices as usual.
The devices have to be turned off before connecting to the charger. If this
is disregarded, the system might get damaged!
Fig. 4.1 shows the pinning of the charge connector. Chargers have to be
connected to pin 2 (+) and pin 4 (GND). An external supply (6…8V, 0.5A)
can be connected to pin 3 (+) and pin 4 (GND).
Use only power supplies
which are certified by mk-messtechnik.
Maximum charging
current is 1 A
Fig. 4.1: Pinning of charge- / buffer connector
Devices must be
turned off before
connecting to
charger, or else the
system might get
damaged!
Pinning of charge- /
buffer connector
Distributed by: Reliant EMC LLC, 3311 Lewis Ave, Signal Hill CA 90755, 4089165750, www.reliantemc.com