54
Irinos IR Users Manual © 2015 - 2016 Messtechnik Sachs GmbH, Germany
Product descriptions
o
The Event
becomes active. Irinos-Boxes with
(Slave) show the event "F24".
o
For Irinos-Boxes with a
(integrated Master), the LED
is turned on permanently.
o
The corresponding hardware status bit of the respective
measurement channel is set. It can be read out by the
measurement software.
Data acquisition
All channels are
sampled synchronously
. It does not matter,
whether the input channel is configured for 1Vpp or TTL / RS422.
Error detection
Channels, which are configured for the input type 1Vpp, have a
channel-wise error detection of the input signals. If everything is ok,
the blue LED is turned on, otherwise it flashes. In parallel, errors can
be detected via the following ways:
o
The corresponding bit in the hardware status of the measurement
channel are set. It can be read out by the measurement software.
o
The event
becomes active. Using the standard settings, it
is not displayed on the
. However, it will be
recorded in the
.
If an error occurs, the position value still can be read out. It
may be valid, but this is not necessarily the case. For a
reliable operation of the measurement application, it is
required to read the
by the measurement
software. In case it contains an error information, appropriate
action regarding the application has to be taken.
4.6
IR-DIO with digital in- and outputs
The Irinos-Box IR-DIO provides 16 digital inputs and 16 digital
outputs. Three types are available:
o
With 4 connectors M23 for the connection of external connection
modules. The supply for the in- and outputs must be supplied via a
M12 connector.
o
With 4 connectors M23 for the connection of external connection
modules. The supply for the in- and outputs is taken from the ILink
supply power (for low power applications).
176
42
41
177
42
182
158
Содержание Irinos IR
Страница 2: ......
Страница 7: ...Introduction...
Страница 15: ...Safety instructions...
Страница 22: ......
Страница 23: ...System overview...
Страница 39: ...Product descriptions...
Страница 64: ......
Страница 65: ...Pin assignment...
Страница 92: ......
Страница 93: ...Mounting Cabling...
Страница 109: ...Setup...
Страница 118: ...118 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Setup...
Страница 120: ......
Страница 121: ...Measurement Control via MscDll...
Страница 168: ......
Страница 169: ...Troubleshooting and first aid...
Страница 186: ...186 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Troubleshooting and first aid...
Страница 187: ...Maintenance cleaning and disposal...
Страница 190: ......
Страница 191: ...Application notes...
Страница 198: ......
Страница 199: ...Specifications dimensions...
Страница 209: ...209 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Specifications dimensions...
Страница 210: ...210 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Specifications dimensions...
Страница 211: ...211 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Specifications dimensions...
Страница 212: ...212 Irinos IR Users Manual 2015 2016 Messtechnik Sachs GmbH Germany Specifications dimensions...