o A significant amount of a diluting dye e.g. methylene blue, indigo green and acid indigo blue or 
carbon monoxide hemoglobin (COHB), methionine (Me + Hb) or hemoglobin, jaundice. 

o Drugs such as dopamine, procaine, prilocaine, lidocaine and butacaine. 

• It is not recommended to use the device if the patient is suffering from carbon mo

noxide 

poisoning. 

 

 

 

4.

 

PRECAUTIONS 

 

 Before use, check the packaging to make sure that the device and accessories fully comply with 

the description and instructions. Otherwise the device may malfunction. 

 Periodically check the main unit and all accessories. Make sure there are no visible damage that 

may impact patient safety and monitor the performance of cables and transducers. It is 
recommended to check the device at least once a week. Discontinue use of the device in the event of 
obvious damage. 

 It is not recommended to use the oximeter in a high electromagnetic frequency environment. 

Do not use the oximeter during MRI and CT examinations. 

 Do not use the device on people allergic to rubber. 

 Do not use the oximeter in an environment with flammable gases such as certain flammable 

anesthetics. Using in these conditions may cause an explosion. 

 Do not disassemble or change the configuration on your own. Except as described in the 

manual 

 SOLUTIONS OF EASY TROUBLE section. 

 

CAUTION:

 

Do  not  repair  the  unit  yourself.  Necessary  repair  and  maintenance  work  must  ONLY  be 
performed by qualified service engineers. Users are not allowed to maintain or repair the device 
themselves.

 

 

 It is recommended that the sensor is not continuously applied to the same finger for more than 

2 hours. Continued use of the device may cause an unpleasant or painful sensation, especially in 
patients with a microcirculation barrier. The device must not be used on swollen, irritated or 
oversensitive skin. 

Содержание MM-155

Страница 1: ...USER MANUAL PULSE OXIMETER MM 155 OXYmed THANK YOU FOR BUYING OUR BRAND PRODUCTS WE WISH YOU A PLEASANT USE BEFORE USING THE DEVICE FOR THE FIRST TIME PLEASE READ THE USER MANUAL ...

Страница 2: ...ers you a pulse oximeter that measures both saturation and heart rate It is extremely simple and convenient to use just put it on your finger and press one button It is a completely non invasive and painless test that anyone can do on their own It is enough for the patient to put one of the fingers into the sensor for diagnosis and the display screen after a few seconds will show the directly meas...

Страница 3: ...he examined person must not have painted nails The subject s nails must not be too long If stable readings cannot be obtained stop measuring The device may not work properly for all patients If the screen displays abnormal indications during the test pull your finger out and reinsert it to restore normal operation Read the measured value when the pulse waveform is normalized and the pulse waveform...

Страница 4: ...nt of obvious damage It is not recommended to use the oximeter in a high electromagnetic frequency environment Do not use the oximeter during MRI and CT examinations Do not use the device on people allergic to rubber Do not use the oximeter in an environment with flammable gases such as certain flammable anesthetics Using in these conditions may cause an explosion Do not disassemble or change the ...

Страница 5: ...o not press the buttons on the front panel with sharp materials The flexible circuit connects the two parts of the device and should not be twisted or pulled Remove the batteries if the device is to be stored for more than a month otherwise leakage may occur The product lanyard is made of anti allergic material if the lanyard causes an allergic reaction stop using it Do not wear a lanyard around y...

Страница 6: ...ences in the optical properties of oxygenated and non oxygenated hemoglobin and calculates the degree of hemoglobin saturation with oxygen SpO2 Rys 1 6 TECHNICAL SPECIFICATION Display TFT Pulse wave display bar graph SpO2 measurement range 0 100 Accuracy 70 100 2 below 70 undefined Optical sensor red light wavelength 660nm 6 65mW infrared wavelength 880nm 6 75mW Heart rate measuring range 30bpm 25...

Страница 7: ...correctly inserting it may damage the device Only use new alkaline or rechargeable batteries in the device Using spent batteries may damage the device and void the warranty Before using the device for the first time or installing new batteries wait about 10 minutes for the device to reach its operating temperature Do not dispose of used batteries in a fire 9 TAKING THE MEASUREMENT 1 Open the clip ...

Страница 8: ...ore cleaning Clean the surface of the device before use First wipe the device with medical alcohol and then allow it to air dry or clean with a dry clean cloth For cleaning use a soft clean cloth dampened with equipment cleaner The device cannot be steam sterilized 11 SOLUTIONS OF EASY TROUBLES SpO2 and pulse rate cannot be displayed normally Place your finger correctly and try again SpO2 and hear...

Страница 9: ...azardous substances entering the environment from electrical and electronic waste Parts of the type BF SpO2 Pulse saturation with oxygen PR bpm Puls rate bmp Serial number Informations about manufacture Date of production Batch number IP 22 IP degree of protection of the device The CE symbol means that this product has undergone an appropriate conformity assessment procedure and meets the essentia...

Страница 10: ...10 Service office 03 948 Warszawa Ul Meksyńska 6 lok 102 E mail mescomptel mescomp pl tel 48 22 838 63 38 opening hours 7 30 15 30 date of actualization 12 03 2020 r ...

Отзывы: