MESA LABORATORIES, INC. DataTrace Division 

 Standard 

Operating 

Procedure 

 

Release Date: August 27, 2010 

Page: 9 of 32 

Revision B 

 

FRB Battery Replacement: 

 
Materials Needed: 

 

 DataTrace

®

 Micropack Tracer FRB Model(s) 

 

Lithium Batteries (Standard Temp: 123599-500; LoTemp: 300041-500) 

 

FRB Maintenance Kit (124717-500) 

 

A supply of clean, dry cloths 

 
The battery change procedure for the FRB model Micropack is simple and straight forward. 
However, care must be taken that these procedures are followed completely to avoid equipment 
damage or personal injury. 
 
We recommend that the disassembly of the DataTrace

®

 Tracers take place in a dry, well-lit 

workstation. The best place for this activity would be close to the location that Tracers are stored 
and programmed and that Tracer batteries are stored and conditioned. 
 

DO NOT USE ANY WRENCHES, PLIERS, VICES, OR OTHER MECHANICAL MEANS TO 
LOOSEN OR TIGHTEN THE FRB TRACER BATTERY COVER. THE TRACER GRIPPER IS 
ADEQUATE TO OPEN AND CLOSE THE BATTERY COMPARTMENT. USING ANY OTHER 
DEVICE MAY DAMAGE THE TRACER AND MAY MAKE REASSEMBLY IMPOSSIBLE.

 

 
 

1.  Clean and dry the Tracer body completely. Make sure no process residue remains on the 

Tracer body that could cause the Tracer to slip from your grasp during disassembly or 
could contaminate the battery compartment during battery replacement. 

 
2.  Grasp the Tracer with the Tracer Gripper in one hand, clasping the base (probe end) 

firmly. With your other hand, using the other Gripper, twist the battery cover 
counterclockwise. 

 
3.  Once loosened, remove the Grippers and the battery cover will unscrew easily. 
 
4.  Remove the old battery by lifting straight out. Do not twist or turn the battery while 

removing. 

 
5.  Dispose of the battery appropriately per local regulations. DO NOT TRY TO 

RECHARGE, DISASSEMBLE, OR INCINERATE THE BATTERY. 

 
6.  Examine the battery compartment for any signs of contamination, pay particular attention 

to the battery socket area. If any contamination is found, remove it, and clean the area 
with a dry cloth. 

 
7.  Examine the threads of both the battery cover and Tracer body for damage and 

cleanliness. If the threads show any damage (e.g., cross threading or “burrs”), do not 
continue with this procedure. Loosely reassemble the Tracer and return it to the factory 
for repair. 
 

7.  Remove the old O-ring. DO NOT USE O-RINGS THAT APPEAR DAMAGED. LEAKS 

COULD RESULT THAT MIGHT DAMAGE THE TRACER AND/OR SHORTEN 
BATTERY LIFE. 

 

 

Содержание DataTrace MPIII

Страница 1: ...ame _____________________________________ Date ______________ Title ____________________________________________________________ Technical Services Name _____________________________________ Date ____...

Страница 2: ...lishing Security Levels 8 Procedures 8 1 System Connection 8 2 Installing DataTrace for Windows Software 8 3 Operating the DTW Program 8 4 Programming Tracers for Study 8 5 Reading Collected Data 8 6...

Страница 3: ...he seal for all Tracers replace and properly lubricate the new O ring whenever the Tracer is opened Ensure that the Tracer cover is properly secured before using Tracers should be kept in their Storag...

Страница 4: ...85 C Note the warning labels on the batteries or battery trays The battery should be replaced by a qualified technician in an anti static and non hazardous environment Disposal of spent batteries mus...

Страница 5: ...2 degrees Fahrenheit relative humidity should stay between 0 to 99 percent non condensing 6 2 PC Interface System MPIII Interface Module 1 The MPIII Interface receives its operating power through the...

Страница 6: ...recording data profiles The greatest risk to a Tracer is being dropped onto a hard surface The sensing probe or the electronics could be damaged by the impact 6 4 Battery Replacement Procedures MPIII...

Страница 7: ...ls out 5 Dispose of the battery appropriately per local regulations DO NOT TRY TO RECHARGE DISASSEMBLE OR INCINERATE THE BATTERY 6 Examine the battery compartment for any signs of contamination and ma...

Страница 8: ...clockwise until resistance is felt It is important not to over tighten the battery cover so a coin or other low torque device should be used in the cover s groove to complete the tightening procedure...

Страница 9: ...body completely Make sure no process residue remains on the Tracer body that could cause the Tracer to slip from your grasp during disassembly or could contaminate the battery compartment during batt...

Страница 10: ...n the Tracer battery compartment has been opened The Tracer is now ready to reassemble with a new battery We strongly recommended that Tracers are NOT left unassembled Damage to components could occur...

Страница 11: ...r to installation in the Tracer The best place for this activity would be close to the location that Tracers are stored and programmed Tracer batteries are replaced and Tracers are assembled and disas...

Страница 12: ...led with security capabilities that include password protection and data encryption When enabled access to various program functions can be strictly controlled In addition the program can enable Elect...

Страница 13: ...f the DTW program 2 The first time DTW is started following installation with an enhanced security level designated the Login screen appears 3 Enter admin as the Login and admin as the Password This i...

Страница 14: ...monly referred to as Electronic Signatures The following functions occur only in Complete security mode 8 Data Signing in DataTrace for Windows requires Tracer reads to be digitally signed by the user...

Страница 15: ...DataTrace for Windows Software DataTrace for Windows includes an automatic installation program that will start the installation when the CD is inserted into the CD ROM drive The DTW installation pro...

Страница 16: ...installation is complete 8 3 Operating the DTW program The computer is the input device to program and retrieve data from the target process via the DataTrace System The DataTrace System software and...

Страница 17: ...another value should be entered in the Run ID up to a maximum of eight characters enter it in the text box Press Enter The Auto Number can be defined by clicking the Auto Number button 2 Change eithe...

Страница 18: ...tab A summary table for the Tracer appears along with the program used to collect the data Below the summary table information is a Comments text box where up to 80 characters of notes can be entered...

Страница 19: ...r screen appears the data report will be printed The printout includes all of the Tracer data displayed on the screen along with the complete data set plus some additional information At the top of th...

Страница 20: ...8 Editing of each of the graph elements is accessible by a right click on the target element An edit menu will appear with various options depending on the nature of the target element Modify anythin...

Страница 21: ...ecommend that temperature points are selected which are critical in reference to the target process For example using Tracers in a sterilization process would suggest using 121 C for a verification po...

Страница 22: ...robe tips in such a way that when they are submerged the tips will be at least 1 away from the bottom or sides of the bath and will be submerged at least 1 below the fluid level Tie wraps can be used...

Страница 23: ...oven or dry well calibrator off CAUTION BE SURE TO WEAR WELDERS GLOVES WITH GAUNTLETS WHEN HANDLING HOT OBJECTS FROM THE OVEN Q After the Tracers have cooled enough to handle comfortably remove them...

Страница 24: ...layed by the Humidity Tracer with the target humidity value from the chamber or the salt solution J If the humidity value indicated by the Humidity Tracer is outside of an acceptable range either retu...

Страница 25: ...rforming field calibrations on DataTrace equipment There are two types of calibrations that are possible The first is for temperature in the Micropack III version temperature only Tracers and the seco...

Страница 26: ...temperature Tracer in either a One Point or a Two Point Calibration cycle The calibration process starts when the Program button is pressed However the calibration screens will not appear until the T...

Страница 27: ...ons displayed 6 On the first screen select the parameter to calibrate and whether a Single Point or a Two Point calibration is needed 7 Select the Tracer Reading what the Tracer is actually reading fr...

Страница 28: ...accuracy the application requires One and Two Point Calibrations optimize the humidity sensor for use in a particular application while a New Sensor Calibration is a special calibration to optimize t...

Страница 29: ...ation is almost the same The primary difference is that for a two point calibration be prepared to expose the target Tracer to two humidity environments during the process and the Calibration Wizard w...

Страница 30: ...ion and will be stored in the Calibration History log 13 Items 10 11 and 12 are repeated for the second point if a Two Point Calibration was selected 14 On the next screen a summary of the one or two...

Страница 31: ...ow point and greater than 80 rH for the high point Ideal new sensor calibration values would be 30 rH and 90 rH Humidity reference values can be obtained from either salt solutions or a humidity gener...

Страница 32: ...Tracers 8 9 1 Viewing Audit Trail The DataTrace for Windows audit trail is accessible through the Report functions Click on the Audit Trail button and the audit trail screen is displayed indicating e...