MESA LABORATORIES, INC. DataTrace Division 

 Standard 

Operating 

Procedure 

 

Release Date: August 27, 2010 

Page: 30 of 32 

Revision B 

 

7.  Retrieve the target Tracer from the reference environment after the required minimum 

stabilization time. Prepare to read the Tracer. 

 

8.  Return to the UTILITIES Tab of the DataTrace Control Panel. Place the Tracer in the PC 

Interface and press Read Tracer. The calibration program will determine that the Tracer 
is in the midst of performing a humidity calibration. 

 

9.  When the Tracer is read following the calibration procedure listed above (steps #1 thru 

#5), answer the questions and fill in the blanks as described in the instructions displayed. 

 

10.  On the first screen, select the parameter to calibrate and whether a Single Point or a Two 

Point calibration is needed. 

 

11. Select the Tracer Reading (what the Tracer is actually reading) from those indicated in 

the drop down box on the left side of the screen. Select the humidity value that is closest 
to the time that the Reference Value was determined. 

 

12.  On the next box, enter a reference value (what the Tracer should be reading), then enter 

a reference name and/or description. Enter a description in the text box on the left side of 
the screen. This data is used to document the calibration and will be stored in the 
Calibration History log. 
 

13.  Items #10, #11, and #12 are repeated for the second point if a Two Point Calibration was 

selected. 

 

14. On the next screen, a summary of the one or two points is displayed showing for each 

point the original and corrected values. This information will be placed into the Calibration 
History report. 

 

15. On the final screen of the calibration, enter any comments in the space provided. Enter 

the Operator’s Name for documentation purposes. Also, to print a Report of Calibration, 
enable that option at this time. 

 
This completes the Two Point humidity calibration process. Press Finish. A Calibration Complete! 
screen will appear, press OK. 
 

REPLACEMENT AND CALIBRATION OF A NEW HUMIDITY SENSOR:

 

This section provides instructions for the physical replacement of the humidity sensor in a 
Humidity Tracer. Following the sensor replacement, the sensor may need to be calibrated. 
Calibration instructions are also included in this section. 
 

DO NOT ALLOW THE HUMIDITY SENSOR TO BE CONTAMINATED BY DIRT, DUST, 
FINGER OILS, SALTS, OR LIQUIDS AS IT COULD DAMAGE THE SENSOR AND/OR IT’S 
CALIBRATION. 
 
WE RECOMMEND THAT POWDER-FREE, LATEX SURGICAL GLOVES ARE WORN AFTER 
THE SENSOR COVER IS REMOVED TO AVOID SENSOR CONTAMINATION. 

 

  The FRB Humidity sensors are shipped with a cover to protect the sensor from 

mechanical shocks and contamination. Do not remove the cover until instructed to do so. 
The MPIII sensors do not need a protective cover. 

Содержание DataTrace MPIII

Страница 1: ...ame _____________________________________ Date ______________ Title ____________________________________________________________ Technical Services Name _____________________________________ Date ____...

Страница 2: ...lishing Security Levels 8 Procedures 8 1 System Connection 8 2 Installing DataTrace for Windows Software 8 3 Operating the DTW Program 8 4 Programming Tracers for Study 8 5 Reading Collected Data 8 6...

Страница 3: ...he seal for all Tracers replace and properly lubricate the new O ring whenever the Tracer is opened Ensure that the Tracer cover is properly secured before using Tracers should be kept in their Storag...

Страница 4: ...85 C Note the warning labels on the batteries or battery trays The battery should be replaced by a qualified technician in an anti static and non hazardous environment Disposal of spent batteries mus...

Страница 5: ...2 degrees Fahrenheit relative humidity should stay between 0 to 99 percent non condensing 6 2 PC Interface System MPIII Interface Module 1 The MPIII Interface receives its operating power through the...

Страница 6: ...recording data profiles The greatest risk to a Tracer is being dropped onto a hard surface The sensing probe or the electronics could be damaged by the impact 6 4 Battery Replacement Procedures MPIII...

Страница 7: ...ls out 5 Dispose of the battery appropriately per local regulations DO NOT TRY TO RECHARGE DISASSEMBLE OR INCINERATE THE BATTERY 6 Examine the battery compartment for any signs of contamination and ma...

Страница 8: ...clockwise until resistance is felt It is important not to over tighten the battery cover so a coin or other low torque device should be used in the cover s groove to complete the tightening procedure...

Страница 9: ...body completely Make sure no process residue remains on the Tracer body that could cause the Tracer to slip from your grasp during disassembly or could contaminate the battery compartment during batt...

Страница 10: ...n the Tracer battery compartment has been opened The Tracer is now ready to reassemble with a new battery We strongly recommended that Tracers are NOT left unassembled Damage to components could occur...

Страница 11: ...r to installation in the Tracer The best place for this activity would be close to the location that Tracers are stored and programmed Tracer batteries are replaced and Tracers are assembled and disas...

Страница 12: ...led with security capabilities that include password protection and data encryption When enabled access to various program functions can be strictly controlled In addition the program can enable Elect...

Страница 13: ...f the DTW program 2 The first time DTW is started following installation with an enhanced security level designated the Login screen appears 3 Enter admin as the Login and admin as the Password This i...

Страница 14: ...monly referred to as Electronic Signatures The following functions occur only in Complete security mode 8 Data Signing in DataTrace for Windows requires Tracer reads to be digitally signed by the user...

Страница 15: ...DataTrace for Windows Software DataTrace for Windows includes an automatic installation program that will start the installation when the CD is inserted into the CD ROM drive The DTW installation pro...

Страница 16: ...installation is complete 8 3 Operating the DTW program The computer is the input device to program and retrieve data from the target process via the DataTrace System The DataTrace System software and...

Страница 17: ...another value should be entered in the Run ID up to a maximum of eight characters enter it in the text box Press Enter The Auto Number can be defined by clicking the Auto Number button 2 Change eithe...

Страница 18: ...tab A summary table for the Tracer appears along with the program used to collect the data Below the summary table information is a Comments text box where up to 80 characters of notes can be entered...

Страница 19: ...r screen appears the data report will be printed The printout includes all of the Tracer data displayed on the screen along with the complete data set plus some additional information At the top of th...

Страница 20: ...8 Editing of each of the graph elements is accessible by a right click on the target element An edit menu will appear with various options depending on the nature of the target element Modify anythin...

Страница 21: ...ecommend that temperature points are selected which are critical in reference to the target process For example using Tracers in a sterilization process would suggest using 121 C for a verification po...

Страница 22: ...robe tips in such a way that when they are submerged the tips will be at least 1 away from the bottom or sides of the bath and will be submerged at least 1 below the fluid level Tie wraps can be used...

Страница 23: ...oven or dry well calibrator off CAUTION BE SURE TO WEAR WELDERS GLOVES WITH GAUNTLETS WHEN HANDLING HOT OBJECTS FROM THE OVEN Q After the Tracers have cooled enough to handle comfortably remove them...

Страница 24: ...layed by the Humidity Tracer with the target humidity value from the chamber or the salt solution J If the humidity value indicated by the Humidity Tracer is outside of an acceptable range either retu...

Страница 25: ...rforming field calibrations on DataTrace equipment There are two types of calibrations that are possible The first is for temperature in the Micropack III version temperature only Tracers and the seco...

Страница 26: ...temperature Tracer in either a One Point or a Two Point Calibration cycle The calibration process starts when the Program button is pressed However the calibration screens will not appear until the T...

Страница 27: ...ons displayed 6 On the first screen select the parameter to calibrate and whether a Single Point or a Two Point calibration is needed 7 Select the Tracer Reading what the Tracer is actually reading fr...

Страница 28: ...accuracy the application requires One and Two Point Calibrations optimize the humidity sensor for use in a particular application while a New Sensor Calibration is a special calibration to optimize t...

Страница 29: ...ation is almost the same The primary difference is that for a two point calibration be prepared to expose the target Tracer to two humidity environments during the process and the Calibration Wizard w...

Страница 30: ...ion and will be stored in the Calibration History log 13 Items 10 11 and 12 are repeated for the second point if a Two Point Calibration was selected 14 On the next screen a summary of the one or two...

Страница 31: ...ow point and greater than 80 rH for the high point Ideal new sensor calibration values would be 30 rH and 90 rH Humidity reference values can be obtained from either salt solutions or a humidity gener...

Страница 32: ...Tracers 8 9 1 Viewing Audit Trail The DataTrace for Windows audit trail is accessible through the Report functions Click on the Audit Trail button and the audit trail screen is displayed indicating e...

Отзывы: