MES SQA-iO Скачать руководство пользователя страница 12

 

Service Manual Version 18

7

.5.1

 

 

AUGUST 3, 2021   SQA-iO Service Manual                     

12

 

 

      

3.

 

Check the capillary after filling (Figure 3), 

visually confirm that the sample 

has completely filled

 the thin section (without a meniscus) and the cuvette 

section.  A small level of sample should appear in the syringe. 

Tap on the 

syringe to make sure there are no air bubbles

 in the sample. If, after 

tapping, some air bubbles appear below the Luer adaptor, dip the capillary into 
the semen sample again and aspirate a small quantity of semen to draw the air 
bubbles into the syringe. 

4.

 

Quickly (to avoid wicking) wipe the outer surface of the capillary 

- both 

top and bottom (Figure 4) with a delicate wipe (Kimwipes, etc.). It is important 
to remove all semen from the exterior of the capillary in order to keep the SQA-
iO clean. Visually confirm that the capillary chambers are still full following the 
cleaning process. If some of the sample has been depleted (meniscus formed in 
the thin part of the capillary) fill the capillary part from the cuvette section by 
slightly pushing in the piston. 

 

Figure 4 Wipe the tip  

 

 

 

Fig 3 Inspect for bubbles 

5.

 

Slowly and carefully 

push-in the blue piston

 until it is level with the plastic 

(Figure 5). 

 

Figure 5:  Push in the blue piston 

6.

 

The capillary is now ready.             

 

 

 

 

 

 

7.

 

Insert the testing capillary into the measurement compartment with the 
blue stopper down.  

Push it in as far as it will go to ensure that the capillary is 

properly seated. 

 

                  

 

 

Содержание SQA-iO

Страница 1: ...Service Manual Version 187 5 1 AUGUST 3 2021 SQA iO Service Manual 1 Service Manual Version 187 5 1 Catalog IO ML 01678 00 Rev August 2021...

Страница 2: ...t 4 USB Port 5 Testing Capillary 5 Maintenance Schedule 5 Testing and Operting Requirements 6 Operating Temperature Humidity Altitude 6 SECTION III Quality Control Internal QC 6 Printing the Service R...

Страница 3: ...ple is withdrawn into an SQA capillary disposable delivery system inserted into the SQA iO where key parameters are collected and sample test results are processed utilizing proprietary technology and...

Страница 4: ...s to extract the required clinical semen parameters Analog to Digital Converter Semen parameters extraction algorithms Self Test and QC KeyPad Microcontroller Operational Monitor 1 2 3 4 Concentration...

Страница 5: ...ng Capillary Sample delivery system Normal sample Requires approximately 0 5cc of semen Short samples Requires approximately 20 l Single use design for testing semen in a biologically safe manner Moti...

Страница 6: ...sing linearly to 50 humidity at 38 C Operates in a wide range of ambient temperatures 15 38 C however the system is calibrated to measure semen samples at room temperature 20 25 C 68 77 F Intended for...

Страница 7: ...nge Instructions for printing the SQA iO SERVICE parameters to prepare for technical support If a Self Test failure occurs the related status icon in the SQA iO HOME SCREEN will turn red Click on the...

Страница 8: ...f the optics you will feel a shelf in the back top section of the chamber Fig 2 and 3 CLEANING STEP 2 1 Use a Fibrous material cleaning paddle Fig 4 supplied in your TEST KIT Moisten with only ONE dro...

Страница 9: ...icon indicating a failure Observe the normal range value 1 Clean the test chamber using ONLY MES cleaning kit 2 Remove and reconnect USB cable 3 Reboot the device If problem persists refer to MES cust...

Страница 10: ...de passed the self test Amplitude failed the self test Suggestion Clean the device ZERO LEVEL 500 525 Zero Level passed the self test Zero Level failed the self test Suggestion Clean the SQA iO Mainta...

Страница 11: ...ting the sample collection container WARNING Do not shake or use a pipette to mix the sample otherwise air bubbles will form and test results will be inaccurate Figure 1 5 Carefully check that the liq...

Страница 12: ...ary both top and bottom Figure 4 with a delicate wipe Kimwipes etc It is important to remove all semen from the exterior of the capillary in order to keep the SQA iO clean Visually confirm that the ca...

Страница 13: ...section while keeping the tip of the capillary well below the sample level and well below the level of any bubbles covering the liquid 3 Withdraw the capillary tip from the semen sample and visually i...

Отзывы: