Service Manual Version 18

7

.5.1

 

 

AUGUST 3, 2021   SQA-iO Service Manual                     

10

 

 

      

 

TROUBLESHOOTING PARAMETERS OUT OF RANGE

: This table describes what to do when a problem occurs with the 

service data key parameters. It is a detailed description of the Self-test failure that occurs on the HOME page. To solve 
the issue, navigate to the SERVICE page, follow the instructions below, and rerun the Self-test. If the problem persists, 
contact customer support. 

PARAMETER 

ACCEPTA

BLE 

RANGE 

DESCRIPTION/SOLUTION 

REFERENCE 1 

(REF 1) 

150 

 350 

mV 

   Reference 1 passed the self-test

 

   Reference 1 failed the self-test 

LED CURRENT 1 

(LED 1) 

 20 mA  

   LED Current 1 passed the self-test

 

   LED Current 1 failed the self-test 

 
Suggestion: Clean the device 

AMPLITUDE 

50 

 100 

mV 

   Amplitude passed the self-test

 

   Amplitude failed the self-test 

 
Suggestion: Clean the device

 

ZERO LEVEL 

500 

 525 

   Zero Level passed the self-test

 

   Zero Level failed the self-test 

 
Suggestion: Clean the SQA-iO.  Maintain ambient temperature between 20 

 25°C / 68 

 77°C 

REFERENCE 2 

(REF 2) 

OPTIMAL: 

2800 

 

3500 

PASSED: 

2500 

 

2800 

   Reference 2 passed the self-test 

   Reference 2 passed the self-test but is not in the optimal    

         range 

   Reference 2 failed the self-test 

 
Suggestion: Clean the device 

LED CURRENT 2 

(LED 2) 

10 

 32 

mA 

   Led Current 2 passed the self-test

 

   Led Current 2 failed the self-test 

AUTO-

CALIBRATION 

AND 

STABILIZATION 

   Auto-Calibration and Stabilization passed (Zero Level parameter is stable)

 

    Auto-Calibration and Stabilization failed 

 
Suggestion:   

 

Remove testing capillary from the measurement compartment 

 

Remove the SQA-iO from sources of vibrations (centrifuge)  

 

Clean the device 

SELF-TEST 

 

   Self-test passed (The key system parameters are in range) 

   Self-test failed  

Содержание SQA-iO

Страница 1: ...Service Manual Version 187 5 1 AUGUST 3 2021 SQA iO Service Manual 1 Service Manual Version 187 5 1 Catalog IO ML 01678 00 Rev August 2021...

Страница 2: ...t 4 USB Port 5 Testing Capillary 5 Maintenance Schedule 5 Testing and Operting Requirements 6 Operating Temperature Humidity Altitude 6 SECTION III Quality Control Internal QC 6 Printing the Service R...

Страница 3: ...ple is withdrawn into an SQA capillary disposable delivery system inserted into the SQA iO where key parameters are collected and sample test results are processed utilizing proprietary technology and...

Страница 4: ...s to extract the required clinical semen parameters Analog to Digital Converter Semen parameters extraction algorithms Self Test and QC KeyPad Microcontroller Operational Monitor 1 2 3 4 Concentration...

Страница 5: ...ng Capillary Sample delivery system Normal sample Requires approximately 0 5cc of semen Short samples Requires approximately 20 l Single use design for testing semen in a biologically safe manner Moti...

Страница 6: ...sing linearly to 50 humidity at 38 C Operates in a wide range of ambient temperatures 15 38 C however the system is calibrated to measure semen samples at room temperature 20 25 C 68 77 F Intended for...

Страница 7: ...nge Instructions for printing the SQA iO SERVICE parameters to prepare for technical support If a Self Test failure occurs the related status icon in the SQA iO HOME SCREEN will turn red Click on the...

Страница 8: ...f the optics you will feel a shelf in the back top section of the chamber Fig 2 and 3 CLEANING STEP 2 1 Use a Fibrous material cleaning paddle Fig 4 supplied in your TEST KIT Moisten with only ONE dro...

Страница 9: ...icon indicating a failure Observe the normal range value 1 Clean the test chamber using ONLY MES cleaning kit 2 Remove and reconnect USB cable 3 Reboot the device If problem persists refer to MES cust...

Страница 10: ...de passed the self test Amplitude failed the self test Suggestion Clean the device ZERO LEVEL 500 525 Zero Level passed the self test Zero Level failed the self test Suggestion Clean the SQA iO Mainta...

Страница 11: ...ting the sample collection container WARNING Do not shake or use a pipette to mix the sample otherwise air bubbles will form and test results will be inaccurate Figure 1 5 Carefully check that the liq...

Страница 12: ...ary both top and bottom Figure 4 with a delicate wipe Kimwipes etc It is important to remove all semen from the exterior of the capillary in order to keep the SQA iO clean Visually confirm that the ca...

Страница 13: ...section while keeping the tip of the capillary well below the sample level and well below the level of any bubbles covering the liquid 3 Withdraw the capillary tip from the semen sample and visually i...

Отзывы: