4.8 Adjustment of column wedges

Remove column casings  (see section 4.3.1 page 15).

Loosen the two retaining screws (1) on either side of the column.

Loosen the two locking nuts (2), one on either side of the column.

Adjust the column so that it moves without exposing gaps

using the adjuster screws (3). 

The adjuster screw torque is 3-4 Nm.

Move the column up and down and check for gaps.

Lock the adjuster screws (3) with locking nuts (2) and ensure that

the adjuster screws remain in place while tightening.

Tighten the retaining screws (1).

4.9 Removal/mounting of back section

Remove motor connector (6)  (see section 4.5.2 page 20)

Remove motor  (see section 4.5.2 page 20)

Loosen screws(1) two turns.

Remove back section by pulling it horizontally

out from middle section.

Loosen frame connector(6) mounting nut(7) 

and remove connector.

Loosen middle section table top screws(2) 

and remove table top(3).

If necessary, remove motor mounting plate(4) 

by loosening screws(5)

Re-assemble in reverse order.

NOTE! Tighten screws (1) to 40 Nm.

If you replace the back section with a different model, 

ensure that the motor is facing the right way.

3

1

2

o1

02

6h

.e

ps

2

1

2

4

5

5

3

7
6

Содержание PRACTICO

Страница 1: ...INTENANCE MANUAL PRACTICO OPERATING TABLE o 1 0 4 7 b p d f Drafted by Markku Luokkanen Type Maintenance manual Completed 15 3 2003 Approved by Arto Koski Laulaja Document DO1046 en Version 2 1 12 200...

Страница 2: ...Weights and dimensions 11 3 3 Structure 12 3 3 1 Main components 12 4 MAINTENANCE AND REPAIR 13 4 1 Maintenance procedures 13 4 1 1 Daily maintenance 13 4 1 2 Monthly maintenance 13 4 1 3 Annual main...

Страница 3: ...val of side tilt motor 22 4 5 6 Mounting of side tilt motor 22 4 5 7 Removal mounting of mattress base longitudinal adjustment motor 22 4 6 Replacement of gas springs 23 4 6 1 Removal of Trendelenburg...

Страница 4: ...4 5 CLEANING 29 5 1Daily cleaning and disinfecting 29 5 1 1 Disinfecting 29 5 1 2 Drying 29 6 RECYCLING 29 6 1Metals and plastics 29 6 1 1 Gas springs 29...

Страница 5: ...manufacturer instructions when using diatherm or defibrillation equipment Combustible anaesthetic gases must not be used with the table Before making any adjustment of the operating table ensure that...

Страница 6: ...ommend recharging the battery regularly overnight after a day s use This ensures that the table is always ready for use and it extends the battery life Charging takes approx 5 10 hours depending on th...

Страница 7: ...tpatient surgical applications The Practico works on both battery and mains power The Practico can be moved from one place to another easily 2 GENERAL Expertise is essential The patient is the most im...

Страница 8: ...erature for at least 6 hours before recharging the battery or switching on to allow any condensation formed to evaporate NOTE All packaging cardboard should be recycled Wood and plastics are energy wa...

Страница 9: ...noperatingtablewithlongitudinaltablesurface adjustment Surface materials Stainless steel accessory rails mattress base parts High pressure laminate base panels Anodised aluminium Polyester hand contro...

Страница 10: ...of shock protection B type equipment Fluid protection Water resistant device IPX4 Cleaning and disinfecting See section 5 page 29 Function type Intermittent operation Combustible anaesthetic gas prot...

Страница 11: ...delenburg 26 Headrest standard 45 30 3 2 5 Weights and dimensions Taulukko 1 Dimensions STANDARD LONGITUDINAL ADJUSTMENT Operating table surface 4 part 4 part Operating table weight 165 kg 175 kg Leng...

Страница 12: ...l point o 1 0 4 5 d p d f Directional castor Control unit Batteries o 1 0 4 6 t p d f Side tilt motor motor Trendelenburg motor Back section Height adjustment Charging outlet 230 120V Switch Power sou...

Страница 13: ...oil Ensure that all castors roll easily and lock securely Check all table functions by fully extending and retracting them Ensure that motor equipped back sections are properly mounted to the seat If...

Страница 14: ...l unit Hand control unit Battery Power source battery If there is a malfunction in one of the following components it usually affects only one function Motor Gas spring Malfunction Cause Table adjustm...

Страница 15: ...onnect power supply First remove the four screws 1 Raise the telescoping column casings and hang from the mattress base Disconnect the hand control connector from the table base Remove casing screws 3...

Страница 16: ...NG Before removing the pin on top of the height adjustment motor 3 ensure that the seat will not fall Push the top of the bellows down and remove the lock ring 1 Remove the pin 2 and raise the seat WA...

Страница 17: ...ack connector Remove the mounting bar by unscrewing 1 Lift out the battery pack 4 4 2 Control unit Disconnect power supply Turn switch to STAND BY position Raise and turn the table base casing see sec...

Страница 18: ...rging charge status controller Disconnect power supply Turn switch to STAND BY position Raise and turn the table base casing see section 4 3 3 page 15 Disconnect the battery pack connector Disconnect...

Страница 19: ...3 3 1 page 12 Open the fuse box and replace fuses 4 4 5 2 Battery fuse Open battery casing cover Replace fuse Carfuse 15A Size 5 3 x 19 1 x 18 7 mm 4 4 5 3 Power source fuses Located inside the power...

Страница 20: ...on 4 3 4 page 15 Remove column casings see section 4 3 5 page 16 Remove support plate 3 Remove top of motor see section 4 3 6 page 16 Remove the lock ring 1 from the bottom of the motor and remove the...

Страница 21: ...x nut 4 Disconnect the motor from the control unit Remove the gas spring 5 loosen the joint screws 6 only until the motor comes free Hold the other gas spring in place to prevent the mattress base fro...

Страница 22: ...the motor until the screw on the other end is in place Tighten the gas spring nut 1 Ensure that the gas spring moves freely 4 5 7 Removal mounting of mattress base longitudinal adjustment motor Remove...

Страница 23: ...tress base support and angle until the lower ends of the gas springs can be fastened Run the motor to its minimum extension and angle the mattress base until the bottom of the motor can be fastened 4...

Страница 24: ...position adjust only by half a turn or one whole turn at a time if the gas spring is still under load turn counterclockwise if lifting the handle does not release the gas spring turn clockwise 4 6 7...

Страница 25: ...edal 1 into its free position Loosen screws 3 Loosen screws 4 Move the brake bar to the side 2 Remove the castor Reinstall castor in reverse order Ensure that the brake pedal and cam positions are in...

Страница 26: ...place while tightening Tighten the retaining screws 1 4 9 Removal mounting of back section Remove motor connector 6 see section 4 5 2 page 20 Remove motor see section 4 5 2 page 20 Loosen screws 1 two...

Страница 27: ...27 4 10 Schematics 1 o 1 0 4 7 c p d f 6 3 5 7 7 8 9 10 11 12 13 14 16 17 18 19 15 2 20 21 1234 4...

Страница 28: ...2 Ah 1 10 71335972 Circuit breaker 1 11 A41527800 Wire bundle Main switch power source 1 12 A41527900 Wire bundle Control unit main switch 1 13 71336074 Control unit 1 14 71335468 T wire 1 15 7133546...

Страница 29: ...8 5 1 1 Disinfecting Wipe down with a 3 chloramine based disinfectant Klorilli or similar cleaning agent Normal excretion stains from patients are cleaned and disinfected with a 10 chloramine based d...

Страница 30: ...t your sales representative For more information on recycling contact your local waste management facility or visit related sites on the Internet Below are recycling symbols which are marked on parts...

Страница 31: ...31 NOTES vaara Meri...

Страница 32: ......

Отзывы: