26
4.8 Adjustment of column wedges
•
Remove column casings (see section 4.3.1 page 15).
•
Loosen the two retaining screws (1) on either side of the column.
•
Loosen the two locking nuts (2), one on either side of the column.
•
Adjust the column so that it moves without exposing gaps
using the adjuster screws (3).
The adjuster screw torque is 3-4 Nm.
•
Move the column up and down and check for gaps.
•
Lock the adjuster screws (3) with locking nuts (2) and ensure that
the adjuster screws remain in place while tightening.
•
Tighten the retaining screws (1).
4.9 Removal/mounting of back section
•
Remove motor connector (6) (see section 4.5.2 page 20)
•
Remove motor (see section 4.5.2 page 20)
•
Loosen screws(1) two turns.
•
Remove back section by pulling it horizontally
out from middle section.
•
Loosen frame connector(6) mounting nut(7)
and remove connector.
•
Loosen middle section table top screws(2)
•
and remove table top(3).
•
If necessary, remove motor mounting plate(4)
by loosening screws(5)
•
Re-assemble in reverse order.
NOTE! Tighten screws (1) to 40 Nm.
•
If you replace the back section with a different model,
ensure that the motor is facing the right way.
3
1
2
o1
02
6h
.e
ps
2
1
2
4
5
5
3
7
6