Appendix II
illumipro-10 Decontamination Record
Decontamination of laboratory equipment must be completed prior to return of the instrument for
servicing. Decontamination must be completed according to the instructions provided. Please
complete the information below and attach the completed record to the instrument prior to return.
Instruments returned for service without evidence of decontamination will not be serviced. If the
instrument is submitted for service without decontamination documentation, it will be shipped
back to the user at the user’s expense.
1. Wipe exterior surfaces of the
illumipro-10
with a lint-free towel moistened with a 10% Bleach-based
Disinfecting Solution. Minimum contact exposure time for kill of Bloodborne Pathogens (Hepatitis-A,
HIV-1, MRSA, SARS, et cetera) is 1 minute.
2. Open
illumipro-10
Lids A and B. Wipe inside lid surface and Heat Block Surfaces with lint-free towel
moistened with 10% Bleach-based Disinfecting Solution. Minimum contact exposure time is 1
minute.
3. Close Lids. Affix the completed
illumipro-10
Decontamination Record to instrument. Package the
instrument in the protective foam packaging and shipping container provided.
Instrument Information
Instrument Serial Number:
Facility Information
Facility Name:
Facility Address:
(Street Address)
(City, State)
Contact Information:
(Name, Title)
(Telephone)
(e-mail)
Decontamination Information
The
illumipro-10
has been decontaminated as described above. Exterior and Interior contact surfaces
have been wiped with a 10% Bleach-based Disinfecting Solution. Minimum contact exposure times have
been monitored.
Performed By:
(Signature/Date)