background image

 

www.avsl.com

                                                600.105UK User Manual

 

 

 

Disconnect circuit power and discharge all 

high-voltage capacitors before testing resistance, 
continuity, diodes or hFE. 

 

Replace the battery as soon as the battery 

indicator 

     

appears. With a low battery, the 

meter may produce false readings that can lead to 

electric shock and personal injury. 

 

Remove the connection between the testing leads 

and the circuit being tested and turn the meter 

power off before opening the meter case. 

 

The internal circuit of the meter should not be 

altered at will to avoid damage of the meter and 

accidents. 

 

A soft cloth and mild detergent should be used to 

clean the surface of the meter on a regular basis. 
No abrasives or solvents should be used to prevent 

the surface of the meter from corrosion, damage 
and accidents. 

 

The meter is suitable for indoor use only. 

 

Turn the meter power off when it is not in use and 

take out the battery when you are not using for a 
long period of time. Check the battery on a regular 

basis as it may leak when it has been out of use 
and replace as soon as any leaks appear to 

prevent damage to the meter.

 

 

The cable tester should not be used on any 

electrified product.

 

 

Do not use the cable tester on any network cable 

where the copper wire connectors are not fully 
pressed to avoid damage to the unit.

 

Содержание 600.105UK

Страница 1: ...Please read this manual thoroughly and ensure all contents are fully understood before use of the apparatus WARNING MTN01 Network Digital Multimeter Item Ref 600 105UK EN61010 1 2010...

Страница 2: ...ltage as marked on the meter between the terminals or between any terminal and grounding The rotary switch should be placed in the right position and no changeover of range should be made during measu...

Страница 3: ...e meter and accidents A soft cloth and mild detergent should be used to clean the surface of the meter on a regular basis No abrasives or solvents should be used to prevent the surface of the meter fr...

Страница 4: ...LCD 3 digits 1999 count 0 5 high Polarity Automatic indicated minus assumed plus Measure method double integral A D switch implement Sampling speed 2 times per second Over load indication 1 is display...

Страница 5: ...www avsl com 600 105UK User Manual Overview COM jack V mA jack 10A jack hFE socket LCD RANGE switch Cable test remote unit...

Страница 6: ...for 1 year 23 5 less than 80 RH DC VOLTAGE OVERLOAD PROTECTION 220V rms AC for 200mV range and 500V DC or 500Vrms for all ranges AC VOLTAGE RESPONSE Average responding calibrated in rms of a sine wav...

Страница 7: ...sounds if resistance is less then 30 20 OVERLOAD PROTECTION 15 seconds maximum 220Vrms DC CURRENT OVERLOAD PROTECTION 500mA 250V fuse 10A range infused MEASURING VOLTAGE DROP 200mV RESISTANCE RANGE R...

Страница 8: ...rn on power of the device or circuit being measured and the voltage value will appear on the Digital Display along with the voltage polarity DC CURRENT MEASUREMENT 1 Red lead to V mA Black lead to COM...

Страница 9: ...est lead to the anode of the diode to be measured and black test lead to cathode 4 The forward voltage drop in mV will be displayed If the diode is reversed the figure 1 will be shown TRANSISTOR hFE M...

Страница 10: ...6 7 8 G RJ45 Remote unit 1 2 3 4 5 6 7 8 G Main unit 1 2 3 4 5 6 RJ11 Remote unit 1 2 3 4 5 6 FAULT DESCRIPTIONS 1 If one cable is open circuit eg no 3 then the no 3 LED on both units wont illuminate...

Страница 11: ...remove the 2 screws in the bottom of the case simply remove the old and replace with a new one Be careful to observe polarity ACCESSORIES Operator s instruction manual Set of test leads red and black...

Отзывы: