2.5 Standard equipment
Electronic fuzzy-supported PID process controller with delayed programme start and programmable
hold time. The controller has permanent power matchinig and an autodiagnostic system for rapid
fault
finding
Recessing push/turn control for simple operation of bath
Visual alarm indication
Mechanical temperature limiter (TB Class 1)
Monitor relay to switch off heating in case of fault
High-grade stainless-steel Pt100-temperature sensor (Mat.Ref. 1.4571) Class A in 4-wire circuit
Boiling mode for temperatures above 95°C
Special equipment (to be ordered separately as accessory): flat cover with concentric ring sets, gable cover
to drain condensate, cooling device, various racks for test tubes, bottles etc
3 Installation
•
•
•
•
•
•
•
The unit must be placed on a
level surface
which cannot tilt.
The vent openings in the left and back side
must
remain unobstructed. Minimum wall spacing on all sides
is 80 mm. The minimum spacing from the top of the bath to the next ceiling (rack, etc.) is 75 cm.
space for
air circula-
tion
space for
air circula-
tion
min.
75cm
min.
80mm
min.
80mm
BASIC