37
Menu mode
5. Set the calibration value to 0.0 K and
accept the setting by pressing the con-
firmation key.
C
K
0,2
-
C
K
C
K
0,2
-
Temperature
Cal1
Cal2
Cal3
37.0
5.0
30.0
0,0
6. Position the sensor of a calibrated
reference instrument centrally in the
appliance's working chamber.
7. Close the door and, in manual mode,
adjust the set temperature to 30 °C.
°C
22
.4
TEMP
°C
30
.0
Set
8. Wait until the appliance reaches the set
temperature and displays 30 °C. The
reference instrument should display
31.6 °C.
°C
30
.0
TEMP
°C
Set 30.0
31.6
°C
9. In the SETUP, adjust the calibration
value CAL2 to +1.6 K (actual value
measured minus setpoint temperature)
and save the setting by pressing the
confirmation key.
C
K
0,2
-
C
K
C
K
0,2
-
Temperature
Cal1
Cal2
Cal3
37.0
5.0
30.0
1,6
+
10. After the calibration procedure, the
temperature measured by the reference
instrument should now also be 30 °C.
°C
30
.0
TEMP
°C
Set 30.0
30,0
°C
With CAL1, a calibration temperature below CAL2 can be programmed accordingly, and
with CAL3, a temperature above. The minimum difference between the CAL values is 10
K.
If all calibration values are set to 0.0 K, the factory calibration settings are restored.