4. Servicing 

4.1. Exchanging the pressing rail _________________________________________________ 

Note: Let the appliance cool down first to avoid 
risks of burns. Then disconnect the appliance 
from the mains power supply. (This is important 
for all repairs, maintenance and servicing work). 

Unscrew the round knob (1).  

Undo the four fastening screws and remove the 
cover of the pressing rail.  

Note: The cutting blade is very sharp! 

The pressing rail (2) is now visible. Remove the 
screw at the right or left end of the rail and taking the 
opposite end of the rail pull it sideways.  

Insert the new pressing rail, screw it in place. When 
replacing the cover make sure that the axis of the 
cutting carriage runs freely along the slit. Finally 
screw the round knob back on. 

 

4.2. Exchanging or turning the cutting blade  _______________________________________ 

Note: Let the appliance cool down first to avoid 
risks of burns. Then disconnect the appliance 
from the mains power supply. (This is important 
for all repairs, maintenance and servicing work). 

Unscrew the round knob (1).  

Undo the four fastening screws and remove the 
cover of the pressing rail. 

Note: The cutting blade is very sharp! 

Turn the blade, or replace it with a new one and 
screw back into position. When replacing the cover 
make sure that the axis of the cutting carriage runs 
freely along the slit. Finally screw the round knob 
back on. 
 

4.3. Spare  parts  ________________________________________________________________ 

Article Name 

 

Article Name 

 

10100 

Sealing rail  

24130 

Signal lamp 

10115 

Pressing rail  

25080 

Spacer (for Art. No. 105, 106) 

12630 Overheating 

protection 

switch 

31190 Heating 

cartridge 

33730 Main 

switch 

17780 

Cutting blade  

19175 Blade 

holder 

 

19635 

Knob for cutting carriage  

59610 Temperature 

regulation 

14405 

Knob for temperature regulation 

 

 

 

Содержание MELAseal 100 Economic

Страница 1: ...zation equipment for medical practice During this period MELAG has succeeded in becoming a leading manufacturer of sterilization equipment More than 335 000 MELAG units sold throughout the world testi...

Страница 2: ...rking with packaging rolls 4 3 Accessories 5 3 1 Work platform 5 3 2 Unit mounted roll holder Comfort 5 3 3 Wall mounted roll holder 5 3 4 Acrylic hood 5 3 5 Standard roll holder 5 4 Servicing 6 4 1 E...

Страница 3: ...adjusted by turning the temperature knob 8 With the transparent packaging material such as MELAfol most commonly used in autoclaves by doctors and dentists one side paper one side foil as described in...

Страница 4: ...ning the lever to the vertical starting position and sealing can be repeated as required if the main switsch is not switch off 2 2 Packaging cutter ____________________________________________________...

Страница 5: ...the MELAseal Package Sealing Device is placed against the wall or if it is mounted directly on the wall then it is best to have a wall mounted roll holder positioned directly above it The wall mounte...

Страница 6: ...blade _______________________________________ Note Let the appliance cool down first to avoid risks of burns Then disconnect the appliance from the mains power supply This is important for all repair...

Страница 7: ...aging which has been opened is not altered if it has been closed again in accordance with regulations A comment on the table in Section 8 2 of the DIN standard says The information in the table only r...

Отзывы: