Chapter 4 - Maintenance 

13 

7. 

Screw on the filter housing lightly with the filter key. 

8. 

Switch on the steam sterilizer and open the shut-off valve at the manometer. 

9. 

Check whether all the parts are secure and that the reverse osmosis unit is free from leaks. 

Replacing the activated carbon filter 

Replace the cartridge of the activated carbon filter once a year or when replacing the fine filter. 

1. 

Switch off the steam sterilizer and close the 
shut-off valve at the manometer. 

2. 

Unscrew the filter housing (a) from the filter 
body (b) using the filter key. Guide the filter 
key upwards from below over the filter 
housing and turn the filter key in a clockwise 
direction. 

 

WARNING 

The filter housing is full to the brim with 
water. Remove the water. 

 

 

 

 

3. 

Remove the activated carbon filter (1) and rinse the filter housing with tap water. 

4. 

Remove the sealing ring from the filter housing (a). 

5. 

Clean and grease the sealing ring lightly with silicone grease (art. no. 24371). 

6. 

Insert the new activated carbon filter in the filter housing. 

7. 

Screw on the filter housing lightly using the filter key. 

8. 

Switch on the steam sterilizer and open the shut-off valve at the manometer. 

9. 

Check whether all the parts are secure and that the reverse osmosis unit is free from leaks. 

Replacing the mixed-bed resin cartridge 

If the mixed-bed resin in the ion exchanger is exhausted (poor permeate quality), replace the mixed-bed 
resin cartridge. Check the permeate conductivity after every replacement of the mixed-bed resin cartridge. 

1. 

Switch off the steam sterilizer and close the 
shut-off valve at the manometer. 

2. 

Unscrew the filter housing (a) from the filter 
body (b) using the filter key. Guide the filter 
key upwards from below over the filter 
housing and turn the filter key in a clockwise 
direction. 

 

WARNING 

The filter housing is full to the brim with 
water. Remove the water. 

 

 

 

3. 

Remove and replace the mixed-bed resin cartridge (1). 

4. 

When inserting a new cartridge, please ensure that the flat seal of the mixed-bed resin cartridge 
(1) points upwards (see arrow). 

5. 

Remove the sealing ring from the filter housing (a). 

6. 

Clean and grease the sealing ring lightly with silicone grease (art. no. 24371). 

7. 

Replace the mixed-bed resin cartridge (1) in the filter housing. 

8. 

Screw on the filter housing lightly using the filter key. 

9. 

Remove the permeate line at the filter housing of the ion exchanger (see Fig. 1, pos. 6). 

Содержание MELAdem 56 M

Страница 1: ...e instrument treatment and hygiene field You our customer are justified in your demand for the best products quality and reliability Providing competence in hygiene and Quality made in Germany we guar...

Страница 2: ...smosis unit MELAdem 56 M About this manual Symbols used Symbol Explanation Draws your attention to a situation which if not avoided could result in damage to the reverse osmosis unit Draws your attent...

Страница 3: ...8 Chapter 3 Commissioning 10 Rinsing the reverse osmosis unit 10 Checking the operational readiness 11 Chapter 4 Maintenance 12 Regular checks and maintenance 12 Replacing the fine filter pre filter...

Страница 4: ...mosis unit please observe the following safety instructions as well as those contained in subsequent chapters Maintenance Use only original replacement parts for maintenance Leaks Close the water inta...

Страница 5: ...s disposed of into the effluent system Mode of functioning An ion exchanger is connected to the reverse osmosis unit to achieve the level of water quality necessary for operation of the Cliniclave 45...

Страница 6: ...Chapter 1 Device description 6 Views of the device Fig 1 Reverse osmosis unit MELAdem 56 M...

Страница 7: ...mosis unit 8 Pre filter Protects the sensitive reverse osmosis membrane The pre filter filter holds back all particulate material rust and other soiling 9 Recessed grip Acts as a carry aid for transpo...

Страница 8: ...ows to connect the reverse osmosis unit 1 Switch off the steam sterilizer at the power switch 2 Remove the caps from the hose connections on the underside of the steam sterilizer and the reverse osmos...

Страница 9: ...unit and connect it to the corresponding connections on the osmosis module 7 Connect the end of the feed water hose to the free screw connection e behind the manometer WARNING The hoses may not be all...

Страница 10: ...unit 4 Place and hold the free end of the feed water line in a 5 l container or similar 5 Switch on the steam sterilizer at the power switch 6 Operate the reverse osmosis unit for c 20 minutes in orde...

Страница 11: ...disappears on the steam sterilizer display a program can be started The reverse osmosis unit now works fully automatically Checking the operational readiness Operating pressure Perform a daily check o...

Страница 12: ...reverse osmosis unit can be found in the maintenance instructions of the steam sterilizer Poor local water quality requires bi annual maintenance of the reverse osmosis unit in accordance with sectio...

Страница 13: ...Switch on the steam sterilizer and open the shut off valve at the manometer 9 Check whether all the parts are secure and that the reverse osmosis unit is free from leaks Replacing the mixed bed resin...

Страница 14: ...k the filter housing 3 and 5 for soiling and rinse out if necessary 7 Insert the filter insert 4 and sealing ring 2 Replace a worn sealing ring 8 Grease the seal face on the lower filter housing 5 wit...

Страница 15: ...S cold water c 1 S cm at 600 S cold water and non exhausted ion exchanger Retention RO Nominal salt retention c 95 Membrane TFC bacteria resistant coil module Pre filter I Particle fine filter 10 5 m...

Страница 16: ...tstra e 6 10 10829 Berlin Germany e mail info melag de Web www melag com Responsible for content Technical Office Subject to technical alterations Your stockist Rev 3 18 1685 Modification date 2018 08...

Отзывы: