Chapter 2 – Installation
9
4.
Connect the waste water line (b
1
) and the
cold water line (c
1
) to the underside of the
steam sterilizer.
5.
Only at Cliniclave 45 M: Place the hoses
over the bracket (d) in the rear area of the
steam sterilizer, so that it protrudes from the
floor unit.
6.
Lead the waste water line (b
2
) and the cold
water line (c
2
) through the duct on the right-
hand side of the osmosis unit and connect it
to the corresponding connections on the
osmosis module.
7.
Connect the end of the feed water hose to
the free screw connection (e) behind the
manometer.
WARNING
The hoses may not be allowed to suffer
kinking or crushing.
8.
Connect the power plug of the reverse
osmosis unit with the power plug of the
steam sterilizer (f) on the underside of the
steam sterilizer and then connect the
bracket.
DANGER
When the steam sterilizer is switched on,
a voltage of 230 V is found on the power
plug of the reverse osmosis unit to the
Cliniclave 45 M/Cliniclave 45 MD.
b
1
c
1
b
2
c
2
e
f
d