2 Installation requirements
9
Water connection
Requirements for the water connection
Feed water
Wastewater
Connection in the
practice
Manual filling via the internal storage
tank.
Optional: automatic via a water
treatment unit e. g. MELAdem 40/ 47
Manual emptying via the internal
storage tank.
Optional: automatically via the one-way
drain with the MELAG upgrade kit for
the tank drain
Wall outlet, nominal width DN 40 or to a
siphon (flush outflow)
Installation height
--
min. 30 cm under the steam sterilizer
Max. water temperature
--
70 °C*
)
Min. flow pressure
Corresponding to the water
treatment unit
--
Recommended flow
pressure**
)
1.5 bar at 3 l/min
--
Min. water pressure
(static)**
)
2 bar
--
Max. water pressure
(static)**
)
10 bar
--
Max. consumption per
program cycle
700 ml
--
Water quality
Distilled or demineralized water in
accordance with DIN EN 13060,
Appendix C
--
*
)
Optional: automatically via the one-way drain with the MELAG upgrade kit for the tank drain
**
)
Optional when using a water treatment unit
With connection to a water treatment unit
MELAdem 40
MELAdem 47
Permissible water
pressure
1.5-10 bar
2-6 bar
Measures for protecting
the drinking water
We recommend the use of a safety combination consisting of a back-flow
preventer and a pipe aerator (e.g. the assembly set in accordance with
EN 1717 for MELAdem) in accordance with EN 1717 to protect against
backflow into the drinking water supply.
Leakage water detector
We recommend the installation of a leakage water detector with a cut-off valve
(e.g. MELAG water stop).
PLEASE NOTE
The outlet hose must be fitted at a constant decline without kinks or sagging. Deviations
to the installation arrangements require consultation with MELAG.
Failure to do so can result in malfunctions of the steam sterilizer.