MTV1000 Ventilator System
56
User’s Manual
6.11.3.
Calibration
Various calibration functions are performed.
Item
Range
Description
Flow Zero
START
/OFF
This function automatically controls offset of inspiratory and
expiratory flow sensor.
O
2
Cell Cal.
OFF
/21%
/100%
Perform calibration of O
2
Cell.
Exp. Flow Auto Cal.
START
/OFF
Perform adjustment of equilibrium for inspiratory flow and
expiratory flow.
EtCO
2
Initial
START
/OFF
Perform calibration of EtCO
2
sensor.
Touch Cal.
ON/OFF
Perform calibration of Touch Screen.
Leak. Test
START
/OFF
Analysis of leakage and compliance is conducted.
CAUTION
When performing O
2
calibration, both 21% and 100% must be carried out.
CAUTION
Before performing Exhalation Flow Auto calibration, inspiration port and expiration port
must be connected to tube.