MTV1000 Ventilator System
29
User’s Manual
4.3.
Oxygen (O
2
) Supply
Oxygen gas can be supplied in two ways. One is using high pressure O
2
port, and the other
is low flow O
2
port.
No.
Figure
Description
1
High Pressure O
2
supply
2
Low flow O
2
supply
WARNING
Since oxygen tank pressure is high, always keep oil away from connector to prevent
explosion.
Before connecting or disconnecting high pressure hose, close oxygen tank valve first.
Check that high Pressure O
2
supply pressure is appropriate (35~90 PSI).
High pressure hose must be kept away from people during use and transportation.
If low flow O
2
supply is not used, keep low flow O
2
port closed up
.