MT-50

 Series     

Problem 

Cause 

Solution 

(6) The eyes feel 

tired easily. The 

right field of view 

doesn’t superpose 

with the left. 

Interpupillary distance is wrong. 

Adjust the interpupillary 

distance.   

Diopter adjustment is wrong. 

Adjust the Diopter. 

The eyepieces for the right are different 

from the left. 

Use the same eyepieces. 

2. Mechanical Part 

(1) Can not get the 

objective focused. 

The cover glass faces down. 

Put  the  cover  glass  to  face 

up. 

The cover glass is not standard.   

Use  a  standard  cover  glass 

with thickness 0.17mm. 

(2) The objective 

touches the cover 

glass while turning 

the nosepiece. 

The cover glass faces down. 

Put  the  cover  glass  to  face 

up. 

The cover glass is not standard.   

Use  a  standard  cover  glass 

with thickness 0.17mm. 

(3) Coarse 

focusing knob is 

too tight. 

Tension knob is too tight. 

Loosen a little. 

(4) Stage declines 

itself.   

Tension knob is too loose. 

Tighten a little. 

(5) Coarse 

focusing knob 

can’t rise. 

The limit stop knob is locked. 

Loosen the knob. 

(6) Coarse 

focusing knob 

can’t decline. 

The base of the condenser is too low. 

Raise the base. 

(7) Can not move 

the slide smoothly. 

The slide is not fixed correctly. 

Adjust it correctly. 

The  movable  specimen  holder  is  not 

fixed properly. 

Adjust it correctly. 

(8) The image 

moves obviously 

when touching the 

stage. 

The stage is fastened incorrectly. 

Fasten the stage correctly. 

3. Electrical Part

 

(1)  The  LED  light 

does not work. 

No power supply. 

Check the connection of the 

power cable. 

The LED light is not inserted correctly.  Insert it correctly. 
The LED lights burnt out. 

Replace it.   

(3)  The  field  of 

view  is  not  bright 

enough.

 

The  use  of  light  adjustment  knob  is 

wrong.   

Adjust correctly. 

(4) The bulb 

flickers or the 

brightness is not 

stable.

 

The wire doesn’t connect all right 

Connect correctly 

 

Содержание MT-50 Series

Страница 1: ...ntenance about MT 50 series biological microscope Please study this manual thoroughly before operating and keep it with the instrument The manufacturer reserves the rights to the modifications by tech...

Страница 2: ...e Diaphragm 12 3 10 Use the Oil Objective 100X 12 3 11 Use the Filter 13 3 12 Replace the Fuse 13 3 13 Assemble and Use of the TV Device 13 3 14 Assemble the Photography Device 14 4 Assembling and Ope...

Страница 3: ...knob or head when moving 4 When working the surface of condenser will be very hot Make sure there is enough room for the heat dissipating around the condenser see Fig 2 5 Connect the microscope to the...

Страница 4: ...lease use these chemical in a ventilated place 2 Don t use organic solution to wipe the surfaces of the other components Please use the neutral detergent if necessary 3 If the microscope is damped by...

Страница 5: ...ries Objective Nosepiece Arm Tension Adjustment Ring Coarse Focusing Knob Fine Focusing Knob Light Adjustment Knob Koehler Illuminator Condenser Binocular Head Eyepiece Objective Stage Koehler Illumin...

Страница 6: ...s Eyepiece Objective Stage Koehler Illuminator Condenser Light Source Coarse Focusing Knob Power Switch Fine Focusing Knob Focus Arm Arm Lock screw Trinocular head Objective Nosepiece Koehler Illumina...

Страница 7: ...tor Filter Block Adjustment knob Light Shield Objective Stage Koehler Illuminator Condenser Koehler Illuminator Condenser Light Source Power Switch Coarse Focusing Knob Objective Nosepiece Arm Focus A...

Страница 8: ...escence Illuminator Eyepiece Objective Nosepiece Arm Stage Fine Focusing Knob Coarse Focusing Knob Light Adjustment Knob Focus Arm Light Shield Objective Koehler Illuminator Condenser Koehler Illumina...

Страница 9: ...lowing is the Assembling Scheme and the numbers denote the assembling order Before assembling make sure there is no dust or dirt Assemble carefully and do not scrap any part or touch the glass surface...

Страница 10: ...6 MT 50 Series LED Fluorescence Microscope Light Source Power Cord LED Fluorescence Illuminator Binocular Head Eyepiece Light Shield Objective Condenser Transformer...

Страница 11: ...oarse focusing knob to lower the stage to a suitable position see Fig 5 2 Install the objectives into the objective nosepiece from the lowest magnification to the highest in a clockwise direction from...

Страница 12: ...LED light source s 2 2 5 Connect the Power Cord Don tuse strong force when the power cord is bended or twisted otherwise it will be damaged 1 Make sure the power switch is at 0 OFF before connecting...

Страница 13: ...ace the Specimen Slide 1 Push the wrench of the specimen holder backwards 2 Loosen the wrench and clamp the slide by the clips while the cover glass faces up see Fig 10 3 Rotate the X and Y axis knob...

Страница 14: ...ig 13 There is 5 diopter on the ring To align with the scale value is your eye s diopter Remember your eye s diopter so that you could use next time 3 6 Adjust the Interpupillary Distance When observe...

Страница 15: ...e center of the field of view see Fig 17 6 Open the field diaphragm gradually If the image is in the center all the time and inscribed to the field of view it shows condenser has been centered correct...

Страница 16: ...0 Use the Oil Objective 100X 1 Use objective 4X to focus the specimen 2 Place a drop of oil on the specimen see Fig 20 3 Rotate the nosepiece counterclockwise and rotate the oil objective 100X to the...

Страница 17: ...see Fig 22 Specification of the fuse T250V 3 15A 3 13 Assemble and Use of the TV Device 1 Loosen the lock screw on the trinocular head and get down the dust cover of the trinocular see Fig 23 2 Get d...

Страница 18: ...aphy device to the trinocular Screw on the lock screw see Fig 24 3 Loosen the lock screw on the photo tube and get down the photo tube see Fig 25 4 Insert the photography eyepiece 3 2X to the eyepiece...

Страница 19: ...the aperture diaphragm indicator to PH position Each magnification phase contrast objective is used with matched ring diaphragm Like 10X phase contrast objective corresponding to 10X ring diaphragm As...

Страница 20: ...center overlap the phase ring center see Fig 27 Fig 28 6 Adjust phase ring and halo of other magnification phase contrast objective according to above steps It can not get the phase contrast microscop...

Страница 21: ...one side of USB data cable into USB output port on the back of observation head see Fig 31 View the microscope video with two eyes by the corresponding video capture and analysis software Digital USB...

Страница 22: ...to the arrow pointed to switch between the transmission bright field and the LED fluorescence Flourescence means flourescence observation and Brightfield means bright field observation See Fig 35 6 I...

Страница 23: ...power supplying The indicator is green in normal while the indicator is orange the indicator becomes green when the pack is fully charged At the same time the last bulb of the light strip 11 is red wh...

Страница 24: ...have accumulated on the lens Clean the lens 4 Unclear image No cover glass on the specimen slide Add the cover glass The cover glass is not standard Use a standard cover glass with thickness 0 17mm Th...

Страница 25: ...Tension knob is too tight Loosen a little 4 Stage declines itself Tension knob is too loose Tighten a little 5 Coarse focusing knob can t rise The limit stop knob is locked Loosen the knob 6 Coarse f...

Отзывы: