IM4830, Page 12 

 

8.2.3 

Depress the two left hand keyboard switches simultaneously to re-initialize the unit.  Confirm the default 
settings:  Frequency = 100 Hz,  Sensitivity = 10.0 and g = 10.0. 

 
 

8.3  

Frequency Accuracy 

 

8.3.1 

Connect the frequency counter to the “mV” output BNC connector.  Set the 4830A sensitivity to 
10.0 and the “g” level to 100.0.  Measure and record the actual frequency output for the following 
displayed frequencies:  1.0, 10.0, 100.0, 1000.0, and 10,000 Hz.  Limits:  ±1%. 
 

8.3.2 

Measure and record the actual frequency output for the following frequencies:  111, 222, 333, 444, 
555, 666, 777, 888, and 999 Hz.  Limits: ±1%. 
 

8.3.3 

Connect the frequency counter to the “TACH” output socket.  Set the 4830A frequency to 100.0 Hz.  
Measure and record the output frequency for the following “Ratio” settings; 1.0, 10.0 and 100.0.  
Limits are ±  1%. 

 
 

Verify “Tach” Output is 5 Volts peak to peak 

 
8.4 

Amplitude Accuracy (mV OUTPUT) 

 

8.4.1 

Connect a true RMS voltmeter and an oscilloscope to the “mV” output socket.  Set the 4830A 
frequency to 100.0 Hz and the sensitivity to 10.0.  Measure and record the mV output for the 
following “g” levels;  limits        ±  1%. 

 

g Level 

mV Output 

Limits 

10.0 

70.7 mVrms 

69.99 – 71.41 

100.0 

707.0 mVrms 

699.9 – 714.1 

1000.0 

7070.0 mVrms 

6999 - 7141 

 

 

Note: Levels < 100 mV pk (70.7 mV Rms) will be affected by noise. 

 
8.4.2  If the measurement results exceed the prescribed limits (see above table), re-adjust  the 

unit using the following method.  Set conditions as per paragraph 8.3.1 and with the “g” 
level set for 100.0.  Adjust RV1 on the analog output board (see fig 3 for the location of 
the potentiometer) ) to obtain 707.0 mVrms at the “mV” output socket. 

 
8.4.3  Re-adjust the 4830A sensitivity to 1.0.  Measure and record the mV output for the 

following g levels; limits 

±

 1%. 

 
 
 

g Level 

mV Output 

Limits 

111.0 

78.49 

77.71 – 79.27 

222.0 

156.98 

155.4 – 158.6 

333.0 

235.47 

233.1 – 237.8 

Содержание ENDEVCO 4830A

Страница 1: ...4830A Accelerometer simulator Instruction manual IM4830A Revision E1...

Страница 2: ...DOCUMENT SUBJECT TO U S EXPORT CONTROLS COMPLIANCE APPROVIAL REQUIRED PRIOR TO DISTRIBUTION IM4830 Page 2...

Страница 3: ...ble Simulator Acceleration Displacement Velocity Tachometer Output signals are provided that simulate voltage mode accelerometers IEPE ISOTRON etc and charge mode accelerometers both single ended and...

Страница 4: ...ED will extinguish The orange trickle charge LED will remain lit indicating that the battery voltage is being maintained topped up by a trickle charge current This configuration is recommended for opt...

Страница 5: ...o increase value or down to decrease value note that the values either side will change when roll over occurs depending on count up or down The cursor has to be returned to right hand area to select a...

Страница 6: ...gnal conditioners The connectors provide the following outputs TACHO Provides a simulated tachometer signal with TTL 5 Volt pulses ICP Provides a millivolt output signal as selected on the panel This...

Страница 7: ...QUIRED PRIOR TO DISTRIBUTION IM4830 Page 7 The rear panel contains the power switch and charger input The LED s labeled F and C indicate fast charge F and trickle charge C The charge mode is selected...

Страница 8: ...DOCUMENT SUBJECT TO U S EXPORT CONTROLS COMPLIANCE APPROVIAL REQUIRED PRIOR TO DISTRIBUTION IM4830 Page 8...

Страница 9: ...DOCUMENT SUBJECT TO U S EXPORT CONTROLS COMPLIANCE APPROVIAL REQUIRED PRIOR TO DISTRIBUTION IM4830 Page 9...

Страница 10: ...DOCUMENT SUBJECT TO U S EXPORT CONTROLS COMPLIANCE APPROVIAL REQUIRED PRIOR TO DISTRIBUTION IM4830 Page 10...

Страница 11: ...the solder joint with shrink tubing 6 Reassemble the case and check for proper operation 8 CALIBRATION AND TESTING Annual calibration is recommended to ensure accuracy and proper performance Calibrati...

Страница 12: ...io settings 1 0 10 0 and 100 0 Limits are 1 Verify Tach Output is 5 Volts peak to peak 8 4 Amplitude Accuracy mV OUTPUT 8 4 1 Connect a true RMS voltmeter and an oscilloscope to the mV output socket S...

Страница 13: ...RT CONTROLS COMPLIANCE APPROVIAL REQUIRED PRIOR TO DISTRIBUTION IM4830 Page 13 444 0 313 96 310 8 317 1 555 0 392 44 388 5 396 4 666 0 470 93 466 2 475 6 777 0 549 42 543 9 554 9 888 0 627 91 621 6 63...

Страница 14: ...TRON ICP Output With the exception of the Tach all other outputs of the 4830A are derived from the mV output To verify their operation suitable calibrated signal conditioners should be used e g for th...

Страница 15: ...8 to 1 02 8 7 Amplitude Accuracy DIFF Charge Output 8 7 1 Connect the DIFF Charge Output to the input of the Model 6634B Signal Conditioner Set the 6634B for Type DIFF Sens 10 00 FS 20 0 Mode g DC pK...

Страница 16: ...ransient power glitches The problem can generally be corrected by doing a Memory Reset as described in item 4 If the problem persists the unit is most likely defective 10 0 TECNICAL SUPPORT For techni...

Отзывы: