Features

n

  Continuous resistance or current measurement on main display.

n

  Optional resistance, current, or figure of merit measurements on  

 

secondary display.

n

  Standard test voltages – 250 V, 500 V, 1 kV, 2.5 kV, 5 kV

n

  Non-standard test voltages – selectable between 50 V and 1 kV in 10 

V  

 

steps, and selectable between 1 kV and 5 kV in 25 

V steps.

n

  Test modes include insulation resistance, polarisation index, step  

 

voltage and dielectric discharge.

n

  The dielectric absorption ratio is automatically calculated if the  

 

corresponding timers are set.  

n

  Either burn or breakdown selectable in insulation resistance mode. 

n

  Insulation ‘Alarm Limit’ available in insulation resistance mode.

n

  Programmable timers include a main test duration timer, plus T1 and  

 T2 timers for time resistance method type tests.

n

  Load capacitance and time constant measurements – displayed at end  

 of test.

n

  LCD backlight.

n

  Data storage, data retrieval and real time data output.

n

  USB or RS232 communications.

n

  Battery level / charge level indicator.

INTRODUCTION 

4

General Description

The S1-552/2 is a microprocessor controlled 5 kV insulation tester 
offering measurement capability of up to 15 T

Ω

. The instrument 

performs automatic tests and has data storage and data retrieval facilities. 

Polarization index and dielectric discharge tests are performed 
automatically, and test duration and voltages can be adjusted according 
to user preference for these tests. A step voltage test can be performed 
automatically, with a default voltage of 1 kV, and test duration of 5 
minutes. It is possible for the user to set different voltages and test 
durations for the step voltage test.

The S1-552/2 can be powered from the mains supply or by its own 
internal rechargeable battery, which provides for at least 6 hours of 
continuous testing with a 100 M

Ω

 load. A battery level indicator on 

the LCD display indicates battery capacity. Connecting power to the 
mains supply connector will automatically charge the battery whether 
the instrument is switched ‘on’ or ‘off’, except during testing. A high 
level of internal isolation allows the instrument to be used while the 
unit is supplied from the mains. An internal battery management system 
switches the instrument off after ten minutes of inactivity. If the battery 
approaches a very low state of charge the instrument turns itself off, and 
mains power must be applied before the instrument can be used again. 
Recorded test results and settings will not be lost when the instrument is 
switched off.

A comprehensive LCD display shows resistance, current, capacitance, 
time constant, voltage, timer measurements, and figure of merit 
measurements such as polarisation index.

Содержание S1-552

Страница 1: ...M S1 552 2 5kV Digital Insulation Tester USER MANUAL...

Страница 2: ...turned off and the circuit discharged manually n Test leads including crocodile clips must be in good order clean and with no broken or cracked insulation n The instrument should not be used if any p...

Страница 3: ...circumstances Failure to follow these instructions could result in damage to the instrument and or a hazard to the operator The word should is used to indicate that the instructions indicate best prac...

Страница 4: ...retrieval facilities Polarization index and dielectric discharge tests are performed automatically and test duration and voltages can be adjusted according to user preference for these tests A step v...

Страница 5: ...e so the ground connection will not be used Power lead connection table Connection UK International USA Earth Ground Yellow Green Green Neutral Blue White Live Line Brown Black If using a fused plug e...

Страница 6: ...dicator Alarm Data recording Test modes Voltage at terminals Battery level Breakdown indicator High voltage warning indicator High voltage warning indicator Analogue display Timers Main T1 T2 Dielectr...

Страница 7: ...isplay will depend upon the test mode selected whether timers T1 and T2 have been set and Power On Off button The instrument will only turn on if this button is pressed held and then released when the...

Страница 8: ...nd current In the IR PI and DD modes the secondary display initially shows the PI polarity index DAR dielectric absorption ratio and on completion of the test the TC time constant and capacitance meas...

Страница 9: ...ns timer defaults to 10 minutes as this is required for a PI reading Figure 3 The result of a dielectric discharge test DD Test settings T1 and T2 times set in order to measure the DAR Test condition...

Страница 10: ...ct the test mode Modes of test to be chosen from include an insulation resistance IR test a polarization index PI test a step voltage SV test and a dielectric discharge DD test The insulation resistan...

Страница 11: ...t now displays clr to indicate clear mode Press and hold down test button until clear process begins shown by analogue display lighting anti clockwise Press the MODE button to exit without deleting th...

Страница 12: ...measured in the terminal Guard terminal For basic insulation tests and where there is little possibility of surface leakage affecting the measurement it is unnecessary to use the guard terminal i e if...

Страница 13: ...s of the bar graph Voltage at terminals The test voltage will default to that selected in the previous test If the instrument has been switched off since the previous test the test voltage defaults to...

Страница 14: ...Increase decrease the decrease the decrease the decrease the test voltage in test voltage in test voltage in test voltage in major steps major steps minor steps minor steps IR only IR only Vs AND Vt...

Страница 15: ...provide accurate measurements with up to 4 mA of noise current present A level of noise exceeding the instruments specified capability will be indicated by an alternating sine wave on the display The...

Страница 16: ...ent symbol PI On completion of the test the instrument will display the insulation resistance recorded at 1 minute and 10 minutes under the segment symbols 1m and 10m 16 TEST MODE SUMMARY Pressing the...

Страница 17: ...ance threshold between the limits of 10 k and 15 T Release the function button to save the current limit Adjusting the alarm threshold level automatically activates the alarm limit mode This is indica...

Страница 18: ...that the test leads are reasonably clean and dry The guard lead can be used to remove the effects of surface leakage if necessary When measuring resistances above 100 G the test leads should not be al...

Страница 19: ...ths may occur Alternatively screened leads are available as an optional accessory from Megger The lead to the negative terminal is fully screened The screen is plugged into the Guard terminal divertin...

Страница 20: ...n test voltage 20 SPECIFICATIONS Capacitance measurement accuracy 23 C 5 5 nF Voltage output accuracy 0 C to 30 C 4 0 10 V of nominal test voltage at 1 G load Current measurement range 0 01 nA to 5 mA...

Страница 21: ...User guide on CD ROM 2000 213 RS232 cable 25955 025 UBS cable 25970 041 Optional Accessories HV test lead sets 3 x 3 m with un insulated small clips 8101 181 3 x 8 m with un insulated small clips 810...

Страница 22: ...Tel 44 0 1304 502 243 Tel 1 610 676 8579 Fax 44 0 1304 207 342 Fax 1 610 676 8625 Megger operate fully traceable calibration and repair facilities ensuring your instrument continues to provide the hi...

Страница 23: ...ry charge level indicator will need to calibrated to the new battery s characteristics End of life disposal WEEE The crossed out wheeled bin placed on the Megger products is a reminder not to dispose...

Страница 24: ...3R 6A8 Canada T 1 416 298 9688 Canada only T 1 416 298 6770 F 1 416 298 0848 E casales megger com Megger products are distributed in 146 countries worldwide This instrument is manufactured in the Unit...

Отзывы: