background image

Press the ‘MODE’ button to cycle through and select the test mode. 
Modes of test to be chosen from include an insulation resistance 
‘IR’ test, a polarization index ‘PI’ test, a step voltage ‘SV’ test, and a 
dielectric discharge ‘DD’ test. If timers T1 and T2 are set, the instrument 
will automatically calculate the dielectric absorption ratio ‘DAR’ of the 
insulation. PI, DAR, and IR values are automatically stored and displayed 
if the data is available, irrespective of the selected test mode. At the end 
of a test the instrument can be made to display either the insulation 
resistances, insulation currents or ratios plus capacitance measurements 
by using the ‘

Ω

/I’ toggle button. See section ‘

Ω

/I button’ for cycling 

through results and toggling units.

Insulation Resistance ‘IR’ test

This is the default mode in which the instrument powers up. This 
test mode measures insulation resistance continuously at the selected 
voltage. The test voltage may be varied during an ‘IR’ test by pressing 
the test voltage 

s

 or 

t

 buttons. The test duration can be set using the 

‘main timer’. The test will finish automatically after this time has elapsed. 
On test completion, insulation capacitance and the time constant 
associated with it is calculated and displayed. 

Time Constant (TC) = R

ins

 x C

ins

 
If timers T1 and T2 are set, the instrument will calculate and display 
the DAR value under the segment ‘DAR’. On completion of the test, 
the instrument will display the insulation resistance measured at these 
times under the segment symbols ‘T1’ and ‘T2’. If the test runs for longer 
than 10 minutes, the instrument will calculate the ‘polarisation index’. 
This value will be displayed under the display segment symbol ‘PI’. 
On completion of the test, the instrument will display the insulation 
resistance recorded at 1 minute and 10 minutes under the segment 
symbols ‘1m’ and ‘10m’.

16

TEST MODE SUMMARY

Pressing the ‘

Ω

/I’ button toggles the display to show the insulation 

resistances, insulation currents, DAR and PI ratios, and capacitance. The 
resistances and currents will be displayed under the ‘T1’, ‘T2’, ‘1m’ and 
‘10m’ segment headings, the ratios under the ‘DAR’ and ‘PI’ segment 
headings. The capacitance reading is indicated by its units of either ‘nF’ 
or ‘µF’. Refer to section ‘

Ω

/I button’.

Breakdown mode / burn mode

The insulation resistance ‘IR’ test operates in either the ‘Burn’ or 
‘Breakdown’ mode of operation. The default mode is breakdown. To 
change mode press and hold the function button, then press and release 
the mode button repeatedly until the desired mode is indicated by the 
flashing symbol on the display. Release the function button to select it.

In the breakdown mode the ‘Breakdown Indicator’ will be illuminated – 
refer to page 6. In this mode the test will automatically terminate should 
the insulation break down to prevent damage to the insulation under 
test.

In the burn mode the ‘Burn Indicator’ is illuminated – refer page 6. 
The burn mode disables the normal breakdown detection and enables 
the insulation test voltage to continue even after breakdown of the 
insulation. This will enable the location of the failure to be seen and 
heard. Due to the potential damage that could occur, the unit produces 
two long beeps when starting a test with burn mode activated.

Alarm limit mode

The insulation resistance ‘IR’ test has an option of setting an alarm limit. If 
this mode is selected, the instrument will beep should the resistance reading 
exceed a user selectable threshold. To set the limit, hold down the function 
button, then press and release the mode button repeatedly until

Содержание S1-552

Страница 1: ...M S1 552 2 5kV Digital Insulation Tester USER MANUAL...

Страница 2: ...turned off and the circuit discharged manually n Test leads including crocodile clips must be in good order clean and with no broken or cracked insulation n The instrument should not be used if any p...

Страница 3: ...circumstances Failure to follow these instructions could result in damage to the instrument and or a hazard to the operator The word should is used to indicate that the instructions indicate best prac...

Страница 4: ...retrieval facilities Polarization index and dielectric discharge tests are performed automatically and test duration and voltages can be adjusted according to user preference for these tests A step v...

Страница 5: ...e so the ground connection will not be used Power lead connection table Connection UK International USA Earth Ground Yellow Green Green Neutral Blue White Live Line Brown Black If using a fused plug e...

Страница 6: ...dicator Alarm Data recording Test modes Voltage at terminals Battery level Breakdown indicator High voltage warning indicator High voltage warning indicator Analogue display Timers Main T1 T2 Dielectr...

Страница 7: ...isplay will depend upon the test mode selected whether timers T1 and T2 have been set and Power On Off button The instrument will only turn on if this button is pressed held and then released when the...

Страница 8: ...nd current In the IR PI and DD modes the secondary display initially shows the PI polarity index DAR dielectric absorption ratio and on completion of the test the TC time constant and capacitance meas...

Страница 9: ...ns timer defaults to 10 minutes as this is required for a PI reading Figure 3 The result of a dielectric discharge test DD Test settings T1 and T2 times set in order to measure the DAR Test condition...

Страница 10: ...ct the test mode Modes of test to be chosen from include an insulation resistance IR test a polarization index PI test a step voltage SV test and a dielectric discharge DD test The insulation resistan...

Страница 11: ...t now displays clr to indicate clear mode Press and hold down test button until clear process begins shown by analogue display lighting anti clockwise Press the MODE button to exit without deleting th...

Страница 12: ...measured in the terminal Guard terminal For basic insulation tests and where there is little possibility of surface leakage affecting the measurement it is unnecessary to use the guard terminal i e if...

Страница 13: ...s of the bar graph Voltage at terminals The test voltage will default to that selected in the previous test If the instrument has been switched off since the previous test the test voltage defaults to...

Страница 14: ...Increase decrease the decrease the decrease the decrease the test voltage in test voltage in test voltage in test voltage in major steps major steps minor steps minor steps IR only IR only Vs AND Vt...

Страница 15: ...provide accurate measurements with up to 4 mA of noise current present A level of noise exceeding the instruments specified capability will be indicated by an alternating sine wave on the display The...

Страница 16: ...ent symbol PI On completion of the test the instrument will display the insulation resistance recorded at 1 minute and 10 minutes under the segment symbols 1m and 10m 16 TEST MODE SUMMARY Pressing the...

Страница 17: ...ance threshold between the limits of 10 k and 15 T Release the function button to save the current limit Adjusting the alarm threshold level automatically activates the alarm limit mode This is indica...

Страница 18: ...that the test leads are reasonably clean and dry The guard lead can be used to remove the effects of surface leakage if necessary When measuring resistances above 100 G the test leads should not be al...

Страница 19: ...ths may occur Alternatively screened leads are available as an optional accessory from Megger The lead to the negative terminal is fully screened The screen is plugged into the Guard terminal divertin...

Страница 20: ...n test voltage 20 SPECIFICATIONS Capacitance measurement accuracy 23 C 5 5 nF Voltage output accuracy 0 C to 30 C 4 0 10 V of nominal test voltage at 1 G load Current measurement range 0 01 nA to 5 mA...

Страница 21: ...User guide on CD ROM 2000 213 RS232 cable 25955 025 UBS cable 25970 041 Optional Accessories HV test lead sets 3 x 3 m with un insulated small clips 8101 181 3 x 8 m with un insulated small clips 810...

Страница 22: ...Tel 44 0 1304 502 243 Tel 1 610 676 8579 Fax 44 0 1304 207 342 Fax 1 610 676 8625 Megger operate fully traceable calibration and repair facilities ensuring your instrument continues to provide the hi...

Страница 23: ...ry charge level indicator will need to calibrated to the new battery s characteristics End of life disposal WEEE The crossed out wheeled bin placed on the Megger products is a reminder not to dispose...

Страница 24: ...3R 6A8 Canada T 1 416 298 9688 Canada only T 1 416 298 6770 F 1 416 298 0848 E casales megger com Megger products are distributed in 146 countries worldwide This instrument is manufactured in the Unit...

Отзывы: