14
display shows the last measurement made until the timer or voltage test
settings are changed, or the test start/stop button is pressed.
Analogue display
This simulates an analogue meter movement to give the user a better
“feel” for how a measurement is progressing. The analogue display
shows resistance only.
The display is also used to indicate how ‘result download’ and ‘deletion
of results’ is progressing.
Secondary display
This part of the display shows the results of ‘time resistance’ method
tests.
‘Pre-Test / During test’ Key action table.
Button press
Key action
Fn + Key action
Pre-test
During test
Pre-test
During test
V
s
/V
t
Increase /
Increase /
Increase /
Increase /
decrease the
decrease the
decrease the
decrease the
test voltage in test voltage in
test voltage in
test voltage in
major steps
major steps
minor steps
minor steps
(IR only)
(IR only)
V
s
AND V
t
Sets voltage
Show set
to 500V
voltage
T
s
/ T
t
Increase /
Display set time Select main
Display time
decrease the
on main timer
timer T1 or T2
set for main
time of the
briefly
timer, T1 or
selected timer
T2 briefly
T
s
AND T
t
Reset timer to
Display set
Reset selected
Display set
zero
time on main
timer to zero
time on main
timer briefly
timer briefly
Mode
Select Test
Cycle through
Mode
IR modes
Ω
/I
Cycle through
Cycle through
Select filter
Select firmware
results display
results display
option
filtering level
of last test
of current
(IR only,
completed
active test
hardware filter
ON)
Record
Turn recording
Select
On/Off
download or
clear
stored
data
V
s
and V
t
represent the test voltage buttons. T
s
and T
t
represent the
timer buttons.