5. 

TESTING AN INSULATION TESTER

Connect the test instrument to the continuity and insulation connec-
tion sockets in the orange and red section of the test box. 

Warning

: Do not apply more than 1250 V d.c. 

5.1 

Voltage load test @ 250 V 

Set the continuity and insulation test select switch to [250 V] and per-
form an insulation test with the test instrument set to 250 V output.
If the test voltage is 250 V d.c. the [V] LED will be on. If the current 
delivered from the unit is greater than 1 mA the [I] LED will also be 
on, showing that the test criteria have been met. A fixed resistor of 
0.25 M

Ω

 +/-3% is used to provide the test reference. The test instru-

ment should indicate a measured value of 0.25 M

Ω

 (+/- operational 

errors including the test instrument’s accuracy).
If either of the LEDs fail to light the test instrument should be re-
turned for a full calibration.

 Record the reading on the record card. 

5.2 

Voltage load test @  500 V 

Set the continuity and insulation test select switch to [500 V] and per-
form an insulation test with the test instrument set to 500 V output.
If the test voltage is 500 V d.c. the [V] LED will be on. If the current 
delivered from the unit is greater than 1 mA the [I] LED will also be 
on, showing that the test criteria have been met. A fixed resistor 
of 0.5 M

Ω

 +/-3% is used to provide the test reference. The tester 

should indicate a measured value of 0.5 M

Ω

 (+/- operational errors 

including the test instrument’s accuracy).
If either of the LEDs fail to light the tester should be returned for a 
full calibration.

 Record the reading on the record card. 

5.3 

Voltage load test @ 1 kV 

Set the continuity and insulation test select switch to [1 kV] and per-
form an insulation test with the test instrument set to 1 kV output.
If the test voltage is 1 kV dc the [V] LED will be on. If the current de-
livered from the unit is greater than 1 mA the [I] LED will also be on, 
showing that the test criteria have been met. A fixed resistor of  

Содержание MTB7671

Страница 1: ...M MTB7671 2 Test box User Guide...

Страница 2: ...Safety Legislation requires them to carry out valid risk assessments of all electrical work so as to identify potential sources of electrical danger and risk of electrical injury such as inadvertent...

Страница 3: ...out supply Fuse Fuse RCD and loop test supply socket Continuity and insulation test select switch Continuity and insulation connection sockets Supply check indicators RCD and loop test select switch...

Страница 4: ...ed and the supply LEDs on the MTB7671 2 will be on To test the RCD plug lightly press the button on the top Test button The red indicator bars should disappear and the supply LEDs on the MTB7671 2 go...

Страница 5: ...this test Record the reading on the record card NOTE If in either of the tests listed above the LEDs fail to light the test instrument under test should be returned for a full calibration 4 3 0 5 Test...

Страница 6: ...reading on the record card 5 2 Voltage load test 500 V Set the continuity and insulation test select switch to 500 V and per form an insulation test with the test instrument set to 500 V output If the...

Страница 7: ...s accuracy Record the reading on the record card 5 5 90 M Test Set the continuity and insulation test select switch to 90 M and perform an insulation check using the tester This connects a 90 M 3 resi...

Страница 8: ...a high current loop test Record the reading on the record card 6 3 Conducting a 180 loop test Set the RCD and loop select switch to Loop 180 and perform a loop test using the tester Check that the lo...

Страница 9: ...I mode and perform the test The test box will trip and disconnect power The Tripped LED will light The trip time should be a nominal 40 ms If the tester fails to indicate a tripped RCD then the tester...

Страница 10: ...ect switch and repeating the tests as described in sections 7 2 to 7 6 NOTE RCD Protection In order to protect the current sense circuit the RCD simulator is protected by a 500 mA fuse If the fuse blo...

Страница 11: ...Pass Fail indication Threshold limits 2 Insulation test voltages 250 V 25 max 500 V 25 max 1 kV 25 max 250 V 25 500 V 25 1 kV 25 Pass Fail indication Threshold limits 2 Insulation test resistances 0 2...

Страница 12: ...2 2 kg EMC In accordance with IEC 61326 1 Operation Safety Meets the requirements of IEC61010 1 Refer to safety warnings supplied Environment Conditions The MTB7671 2 is intended for indoor use only C...

Страница 13: ...is necessary to retun an instrument for repair a Returns Authorisation number must first be obtained by contacting one of the addresses shown You will be asked to provide key information such as the i...

Страница 14: ...SW 2125 Australia T 61 0 2 9659 2005 F 61 0 2 9659 2201 E ausales megger com Megger Limited 110 Milner Avenue Unit 1 Scarborough Ontario M1S 3R2 Canada T 1 416 298 9688 Canada only T 1 416 298 6770 F...