APPLICATION NOTE

 

 

RCD impedance uplift elimination. 

Megger Ltd, Dover, United Kingdom  

www.megger.com 

Sales office: 01304 502 101,  

[email protected]

   

RCD impedance uplift elimination 

Technical support: 01304 502 120,  

[email protected]

  

P3 

Higher currents reduce coil impedance but will cause the RCD to trip, preventing the test from completing. 

The impedance of the coil (A) at 15 

mA is 0.5 Ω.  This value is added to the basic circuit impedance and 

the total is displayed on the MFT.  The MFT cannot distinguish between the impedance of the circuit and 

that of the coil.  

It  is  irrelevant  how  much  current  is flowing  in the  L-N circuit.   The  example  in  Figure 3  shows  the  test 

process in a more realistic circuit, with a motor load and a loop impedance test. 

 

Figure 3.  Test process with motor load and loop impedance test. 

Still,  only  the  difference  of  15  mA  is  generated,  the  majority  returning  down  the  neutral  conductor.  

However this 15 mA difference is the cause of the additional impedance.  

Not all RCDs or RCBOs affect the loop impedance value.  However as it is not possible to distinguish 

which of those devices are suspect the problem can be very confusing, annoying and time consuming. 

The solution: 

The  new  MFT1845  3-wire  loop  impedance  test  uses  a  combination  of  test  currents  and  measurement 

methods to ensure the impedance of coil (A) is not added to the circuit impedance.  

By developing a range of tests within the 3-wire loop test, not only is the method immune to 30 mA RCDs 

and RCBOs, the method will also work for 10 mA devices.  

The new MFT can be identified by the number MFT1845 for European applications. 