Downloading results to a PC:

1. Run Megger Powersuite Professional or Megger Download Manager on 

the PC

2. Select the appropriate downloading instrument from the list

3.   Select "Download from Tester"

4.  Connect the tester to the PC using the USB test lead.

5. Set the tester range knob to [Snd].

6.  The test data will automatically download its contents to the PC.

A bar graph shows the status of the download.

Deleting test results

To delete the latest test result:

1. Set the range knob to [dEL]. The display will flash "dEL" followed by the

test number to be deleted.

2. Press the OK button. The last test result will be will be deleted.

WARNING:

This operation is not reversible. 

To delete all test data:

1.  Set the range knob to [dEL]. The display will flash "dEL".

2. Press the NEXT or LAST key. The display will flash "ALL".

3. Press the OK button. All the test results will be will be deleted.

A bar graph shows the progress of the deletion. 

WARNING:

This operation is not reversible. All data will be deleted. 

18

REPLACING BATTERIES

Batteries

Battery type: 

8 x LR6 (AA), 1.5 V Alkaline, or 8 x 1.2V NiCAD, or 8 x 1.2V

NiMH

Low battery warning symbol

The battery condition is continuously displayed by the symbol 

When the batteries are exhausted, symbol will show 

and the

isntrument switches off. Batteries should be replaced when 2 bars are

displayed.

If the symbol appears as less than fully charged with new batteries fitted,

check for correct polarity.

Note:

Fully charged NiMH or NiCAD rechargeable batteries show a lower

charge than Alkaline batteries, and may not give much warning before

becoming exhausted.

To replace batteries

Warning: Do not switch the instrument on with the battery cover removed.

1. Switch off the instrument and disconnect (the instrument) from any

electrical circuits.

2. The rear cover must not be opened if the test leads are connected. 

3. To remove the rear cover release the screw at the bottom of the cover

and lift the cover upwards.

4. Fit new batteries observing the correct polarity as marked on the battery

compartment.

5. Replace the cover.

Содержание LT310

Страница 1: ...M LT300 Series Earth Loop Testers USER MANUAL...

Страница 2: ...orities may recommend the use of fused test leads when measuring voltage on high energy systems The battery cover must be in place whilst conducting tests Voltage indicator LED s cannot reveal a N PE...

Страница 3: ...warnings 3 Introduction 4 General description 5 Case contents 7 LCD display 6 Front panel 6 Connection panel 7 Lid open closure 7 Preparations for use 8 Batteries 8 Preliminary test lead check 8 Gener...

Страница 4: ...test Phase Phase loop test PFC display INTRODUCTION GENERAL DESCRIPTION 4 Thank you for purchasing the Megger Earth Loop Impedance Tester For your own safety and to get the maximum benefit from your...

Страница 5: ...s 1 x Plug ended test lead 8 x AA LR6 batteries fitted in instrument 1 x Warranty card 1 x Certificate of test 1 x Calibration certificate 1 x CD containing user manual 1 x Quick start guide 1 x Downl...

Страница 6: ...6 LCD DISPLAY FRONT PANEL...

Страница 7: ...any time OK Final operation to save the result Selects the job db circuit number ie 01 02 03 etc Lid open closure 1 Open lid by lifting up front panel tab 1 2 Fold away underneath instrument 2 3 and p...

Страница 8: ...strument will warn the operator and prevent further testing until the instrument has had a chance to cool down Fuse Blown A fuse blown will prevent the instrument from making further tests The fuse in...

Страница 9: ...hot Indicates the instrument needs to cool down before it can continue loop testing Noise on the circuit under test may affect the reading Setup procedure Reverse polarity or line neutral swapping Thi...

Страница 10: ...appliance The mains plug test lead supplied with the Megger tester is a test lead that forms part of the measuring circuit of the instrument The overall length of this lead must not be altered If the...

Страница 11: ...ipping loop test No Trip Earth loop impedance measurement at a power socket The No Trip range is a high resolution 0 01 low test current earth loop resistance measurement range It requires a connectio...

Страница 12: ...ssing the TEST button 12 Bonded metalwork testing Repeat the above test but with the green lead connected to the exposed metalwork For a Hi current phase to earth loop impedance measurement at a power...

Страница 13: ...display will show 480 V Over temperature hot To protect the instrument from over heating during Loop testing thermal protection is fitted If the message hot appears in the display together with the G...

Страница 14: ...d result must be stored in the instrument s memory Zref prior to measuring R1 R2 Saving a Zref distribution result 1 Switch the range knob to Zref 2 Select a No Trip or Hi loop test as appropriate 3 C...

Страница 15: ...the phase to earth voltage Phase to Phase voltage measurement LT320 and LT330 NOTE measured voltage must not exceed 300 V phase to earth To measure the voltage of the electrical supply 1 Set the inst...

Страница 16: ...show the phase sequence indicates R B Y sequence or 1 2 3 indicates R Y B sequence or 1 3 2 NOTE If one of the lines is faulty neither of the symbols is displayed and just the normal neon polarity ind...

Страница 17: ...r is displayed 3 Press OK and the test result will be displayed Note Only the last test result can be recalled to the display To recall the PFC value press LAST or NEXT value if applicable 17 TEST RES...

Страница 18: ...NING This operation is not reversible All data will be deleted 18 REPLACING BATTERIES Batteries Battery type 8 x LR6 AA 1 5 V Alkaline or 8 x 1 2V NiCAD or 8 x 1 2V NiMH Low battery warning symbol The...

Страница 19: ...tended periods of time Fuse Blown indication The fuse blown symbol findicates that an internal fuse has failed This instrument is fitted with a factory fitted fuse and should only be replaced by an au...

Страница 20: ...rement Range 0 V to 500 V Accuracy 2 2 digits Frequency measurement LT320 and LT330 Range 25Hz to 450Hz Accuracy 25 0Hz to 199 9Hz 0 1Hz 200Hz to 450Hz 1Hz Phase rotation indicator LT320 and LT330 Thr...

Страница 21: ...idity Operating range 5 C to 40 C Operating humidity 93 R H at 40 C max Storage range 25 C to 70 C Maximum altitude 2000 m Environmental protection Weather proof to IP54 Safety Meets the requirements...

Страница 22: ...ng amendment No 1 Basic and service errors Loop test ranges The basic error is the maximum inaccuracy of the instrument under ideal conditions whereas the service error is the maximum inaccuracy takin...

Страница 23: ...plug test lead BS 1363 BS Versions 6220 740 Mains plug test lead CEE 7 7 EN versions 6220 741 Mains plug test lead AS NZS 3112 AU versions 6220 790 USB download lead 25970 041 Download Manager CD 6111...

Страница 24: ...ng and Instrument for Repair If it is necessary to retun an instrument for repair a Returns Authorisation number must first be obtained by contacting one of the addresses shown You will be asked to pr...

Страница 25: ...er com Megger Pty Limited Unit 26 9 Hudson Avenue Castle Hill Sydney NSW 2125 Australia T 61 0 2 9659 2005 F 61 0 2 9659 2201 E ausales megger com Megger Limited 110 Milner Avenue Unit 1 Scarborough O...

Отзывы: