www.megger.com
DLRO10HD and DLRO10HDX
25
Maintenance
Maintenance
Routine Inspection
Look for any cracks or other damage to the enclosure, missing ports, etc.
Cleaning
Disconnect the instrument from the Mains Power. Wipe it with a clean cloth slightly damped with water or
Isopropyl alcohol (IPA). Care should be taken near the terminals, the IEC Mains Power and USB sockets.
Allow the instrument to completely dry before it is used.
Instrument Care
The instrument should always be handled with care and not dropped. Always make sure that the instrument is
secured when being transported to prevent mechanical shock.
Test Leads
Leads are silicone insulated and work well in all weather conditions. Always keep the leads in a suitable lead bag
when in storage or transportation.
Regular inspection of leads is recommended to make sure that they are not damaged in any way. Damaged leads
could affect resistance readings and are a safety hazard.
Mains Power Fuse
Always use the correct rated fuse (see
Specifications (page 26)
).
Battery Care
Caution: Batteries are only to be installed or removed by an Authorised Service Centre. Do not attempt
to remove the batteries from this instrument.
To prevent deep battery discharge, the battery should be charged at a minimum of three month intervals
Never attempt to charge the battery below 0 °C (32°F) or above 40 °C (104 °F) ambient
To improve battery life, store the instrument in a cool, dry location
Battery Charge
The battery is charged when Mains Power is connected (unless a test is in progress).
For optimum battery life, charge the battery after each use. A fully discharged battery takes eight hours to
recharge.
The charge level will go from low to full charge in incremental steps and continue as long as Mains Power is
connected (unless a test is in progress). When the battery is fully charged the battery icon will stay steady.
Battery full charge
Battery charge low
Battery discharged: The instrument shuts down automatically.