DLRO100E, 100X & 100H – 100A

3

G

 Safety warnings

 

These must be read and understood before use. Retain these safety warnings for future reference.

 

CAUTION: 

THE INSTRUMENT MUST BE OPERATED ONLY BY SUITABLY TRAINED AND COMPETENT PERSONS

 

Users of this equipment and their employers are required by National Health and Safety Legislation to carry out valid 

risk assessments of all electrical work so as to identify potential sources of electrical danger and risk of electrical injury. 

 

The instrument must NOT be used if any part of it is damaged.

 

Damaged test leads must NOT be used. Test leads, connectors and mechanical guards must be in good order, clean and 

with no broken or cracked insulation.

 

Fan filters and covers giving access to internal conductive parts must be correctly fitted before use.

 

Testing inductive circuits can be hazardous: The DLRO100 is a high power instrument, designed for testing resistive 

loads. It must NOT be used to test inductive loads.

 

DANGER!

 The instrument is not fully protected when switched off.

•  Switch the instrument ON before connecting to the test subject.
•  The test subject must be switched off, de-energised and checked before test connections are made. Ensure that the 

test subject cannot be re-energised whilst the instrument is connected.

•  Do not leave the equipment unattended when connected to the test subject.
•  Do not leave the equipment connected to the test subject after the test is completed.

 

The user must exercise caution when connecting to and disconnecting from the test subject.
•  Always connect test leads to the instrument before attaching to test subject.
•  Keep hands behind tactile barriers on probe clips and clamps when making or breaking connections.
•  High current connections between the instrument and test subject must be secured against accidental detachment 

and must not be disengaged whilst test current is flowing.

•  High current test leads must be a minimum of 3 milliohm resistance.
•  Circuit terminals must not be touched during test.
•  Do not disconnect the instrument from the test subject until the test current has stopped and the TEST warning 

indicator is extinguished.

•  Test leads and connections may become hot during use. Exercise caution when handling.
•  Disconnect from the test subject before switching the instrument OFF.

 

DANGER!

 Some models can be operated by remote control. A test can be started at any time by remote control. These 

additional precautions must be taken for instruments with a remote control function.
•  Measurement connections must be handled only when precautions have been taken to prevent a remote control 

test start.

•  If the remote control link should fail, the test must be stopped manually by pressing the TEST button.

 

If this equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may 

be impaired.

 

Fan filters are user-serviceable. Disconnect all measurement leads and switch the instrument OFF before servicing the 

fan filters.

 

There are no user-serviceable parts inside the instrument; all servicing, including fuse replacement, must be referred to 

Megger approved service centres.

 

Use only Megger approved battery pack, and follow the instructions provided with the battery

 

Warning!

 This instrument contains a Lithium Ion High Energy Battery Pack and a lithium coin cell.

•  Do not pierce, damage, disassemble or modify the battery. The battery contains safety and protection devices 

which, if tampered with may cause the battery to generate heat, rupture or ignite. 

•  Never heat (or dispose of) the battery in a fire.
•  Do not subject the battery to strong impact, mechanical shock or excessive heat.
•  Do not expose the battery to water, salt water or other liquids, or allow the battery to get wet.
•  Never short-circuit or reverse the polarity of the battery pack.
•  In the event of a battery cell leaking, do not allow the released fluid to come into contact with the skin or eyes. If 

contact has been made, wash the affected area with plenty of water and seek medical advice immediately.

Содержание DLRO100E

Страница 1: ...DLRO100E 100EB 100X 100XB 100H 100HB 100A High Performance Digital Low Resistance Ohmmeter User guide...

Страница 2: ...1 Testing modes 10 2 Instrument set up 12 2 DLRO100A Resetting Battery Charge 16 3 Saving a test record 17 4 Delete results 18 5 Download a test record 19 6 Remote operation 20 7 Asset tagging 21 8 S...

Страница 3: ...s must be a minimum of 3 milliohm resistance Circuit terminals must not be touched during test Do not disconnect the instrument from the test subject until the test current has stopped and the TEST wa...

Страница 4: ...ne to earth voltage at which it is safe to connect CAT IV Measurement category IV Equipment connected between the origin of the low voltage mains supply and the distribution panel CAT III Measurement...

Страница 5: ...and test duration for flexibility Battery capacity 200 single tests or up to 2x 10 minutes continuous 100 A output for extended use Mains only units for manufacturing and production applications where...

Страница 6: ...tton 100X and 100H only 8 Range rotary switch 9 TEST button with associated warning lamp 10 Navigation OK buttons 11 Test mode rotary switch 12 Backlight button 13 Display 14 LED indicating line power...

Страница 7: ...harging will be prevented Functional verification Simply turning on the instrument at the test mode switch will initiate a start up process and the display will respond The initialisation screen right...

Страница 8: ...retrieve records 100X and 100H Only A spanner enabling instrument and test settings A custom test selection 10 A 50 A and 100 A pre set test currents Save button 100X and 100H only Manual Auto Contin...

Страница 9: ...s inhibited Disconnect from the earth terminal when this feature is not in use Section of test piece under test The current terminals C1 and C2 must be connected outside of the potential terminals P1...

Страница 10: ...button to start a custom test To configure test setting for user configured test date and time for saved results refer to Instrument set up section 2 6 1 4 User configured Manual and Auto test Progre...

Страница 11: ...2 6 1 10 Test aborted by instrument Aborted test A number of conditions can cause an abort This example shows a bad connection 1 11 Test stopped by user User stopped test by pressing the test button 1...

Страница 12: ...s the current setting is limited to a maximum of 59A OK OK 2 4 Set test duration Test duration is adjusted with UP and DOWN arrows Minutes and seconds are selected with LEFT and RIGHT arrows Select OK...

Страница 13: ...elect OK to accept 2 10 Set Time Adjust hours with UP and DOWN arrows Navigate using LEFT and RIGHT arrows or select OK to accept 2 11 Set Hours 100A 50A 10A 100A 50A 10A 100A 50A 10A 100A 50A 10A OK...

Страница 14: ...LEFT and RIGHT arrows or select OK to accept 2 14 Bluetooth Pairing Pair instrument from PC Smart Device Enter PIN 1234 on PC 2 13 Bluetooth Begin pair Press and hold OK for 3 seconds to begin pairin...

Страница 15: ...0 00 mV A Navigate using LEFT and RIGHT arrows or select OK to accept NOTE The clamp must be zeroed before measurements start for accurate readings The clamp current measurement is not displayed on th...

Страница 16: ...Progress to Battery Reset screen AC is shown if charger lead plugged in Remove lead to proceed 2 21 Lead removed Press Up key to proceed Press Ok to proceed Reapply the AC Menu automatically proceeds...

Страница 17: ...3 1 End of test 3 2 Completion of save Press the SAVE button to save results Save completed Showing date time and slot number for 2 seconds then reverts to the End of Test screen 100A 50A 10A MAN AUTO...

Страница 18: ...all auto test results ii Continuous test log results every 5 seconds iii If internal memory is full delete some records to create space refer to Saving a test record section 3 3 100A 50A 10A MAN AUTO...

Страница 19: ...ults 4 1 Delete results Single 4 2 Delete results Progress screen Delete last saved slot Press OK to accept Press OK to accept OK OK OK 4 3 Results screen Delete ALL Delete all results Select LEFT or...

Страница 20: ...countdown counter will show number of records being downloaded 5 Download a test record Recall a test record 100A 50A 10A 100A 50A 10A 5 3 Listing slot numbers 5 4 Test record Press OK to drill into...

Страница 21: ...be remotely saved on the instrument The user can save results into a Power DB form on the PC Remote control can be terminated from the PC or by moving either rotary switch If a test is in progress the...

Страница 22: ...download Moving the rotary switch to the PC position will show the download screen Test switch can be MAN AUTO or CONTINUOUS For Bluetooth pair as defined before here Tagged data received successfully...

Страница 23: ...lliOhms 5 W MicroOhms 1 2 3 4 5 6 7 8 9 10 11 12 1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 1 2 3 4 1 Delete 2 Download Remote operation 3 Retrieve results 4 Settings 5 Save 6 DualGround Clamp 7 USB 8 Bluetooth 9...

Страница 24: ...the internal temperature of the instrument exceeds a safe level the test will be aborted and indicated on the screen The temperature must drop before testing can be continued Fully charged battery 50...

Страница 25: ...ould be charged at a minimum of 3 month intervals This is to prevent deep discharge Never attempt to charge the battery below 0 C or above 40 C ambient The battery is charged by connecting line power...

Страница 26: ...k Battery charge time Full charge in 2 5 Hrs from flat Maximum output voltage 2 V battery 3 V AC EMC IEC61326 1 Safety IEC61010 CAT IV 600 VAC 500 VDC up to 2000 m CAT IV 300 V up to 4000 m Dimensions...

Страница 27: ...ium Ion Battery Pack 1005 973 Acknowledgments The DLRO100 i uses the FreeRTOS operating system from http www freertos org ii contains RADSOK technology from Amphenol RADSOK is a registered trademark o...

Страница 28: ...ave been defective as claimed Any unauthorised prior repair or adjustment will invalidate the warranty Misuse of the instrument from connection to excessive voltages fitting incorrect fuses or by othe...

Страница 29: ...10 676 8610 MEGGER INDIA PVT LIMITED 211 CRYSTAL PARADISE MALL OFF VEERA DESAI ROAD ANDHERI W MUMBAI 400 053 INDIA T 91 22 2674 0468 F 91 22 2674 0465 MEGGER GMBH OBERE ZEIL 2 61440 OBERURSEL GERMANY...

Отзывы: