background image

Figure 4: schematic for three-terminal resistance measurement with lead null

 

In this diagram, the C1 and P1 terminals are connected together at the electrode under test. This is the “three-terminal with 
lead null” configuration which is only applicable in four-terminal testers. This configuration allows the resistance of the P1 
lead to the electrode under test to be “nulled” out. For three-terminal testers or when lead null is not required, only the P1 
terminal (or X terminal on a three-terminal instrument) connects to the electrode under test. This is shown in Figure 5.

Figure 5: schematic for three-terminal resistance measurement without lead null

 

Principle of operation (three-terminal resistance measurement using ART)

The classic three-terminal test method has a disadvantage, namely that the electrode under test must be disconnected from 
the system it is supposed to protect in the event of a power system fault. The reason for this is that the injected test current 
will take all possible routes to ground and not all of it will necessarily flow through the electrode under test. In this case, the 
instrument will make a reading of the entire earthing network, not just the individual electrode.
By using a current transducer (the Megger MCC CLAMP) to measure the current flowing through the electrode under test as 
a fraction of the total test current injected, the instrument can determine the individual resistance. This arrangement is shown 
in Figure 6.

Figure 6: schematic for three-terminal resistance measurement using ART without lead null

 

In this configuration, the injected test current I splits along two paths into I1 (flowing into the connected earthing system) and 
I2 (flowing into the electrode under test, i.e. I=I1+I2. The resistance of the electrode under test is calculated as R=V/I2 or R=V/ 
(I-I1). The current transducer (MCC CLAMP) measures I2 and feeds this value back to the instrument. 

MCC

Содержание DET Series

Страница 1: ...DET Series EARTH GROUND ELECTRODE TESTERS USER GUIDE...

Страница 2: ...d throughout by Double Insulation c Equipment complies with current EU directives Equipment complies with C tick requirements Do not dispose of in normal waste stream CAT IV 100 V Over voltage categor...

Страница 3: ...instruments This user manual describes the operation and functions of the DET series of Earth Ground Testers DET3TA DET3TC DET3TD DET4TD2 DET4TR2 DET4TC2 DET4TCR2 MCC CLAMP MVC CLAMP CALIBRATION ADAP...

Страница 4: ...t from the earthing system under test while batteries are being changed or the fuse replaced Always replace batteries and fuses with parts of the correct type and rating Do not replace the rechargeabl...

Страница 5: ...als and the instrument should be switched off Warning To avoid damage by leaking electrolyte or deep discharge do not leave cells fitted in an instrument which will remain unused for extended periods...

Страница 6: ...wire measurement and offer a digital display The DET4TR2 and DET4TCR2 provide a rechargeable battery power source The DET4TC2 and DET4TCR2 offer variable frequency testing 94 Hz 105 Hz 111 Hz and 128...

Страница 7: ...ent n n n n n n n Current meter n n n LCD display n n n n n n Backlit display n n Moving coil meter n IP54 rated n n n n n n n EN61010 1 100 V CAT IV n n n n n n n Built in battery charger n n Each in...

Страница 8: ...usted refer to the section on battery replacement or recharging Warning Do not switch the instrument on with the battery cover removed Inspection Before each use of the instrument visually inspect the...

Страница 9: ...he the 25 V 50 V button The procedure for changing the output voltage is as per the DET3TA Auto power down To extend battery life the instrument will automatically switch off six minutes after the las...

Страница 10: ...MP current DET3TC DET4TC2 and DET4TCR2 only Display symbols DET3TA Symbol Meaning Rc Current stake C Stake resistance exceeds range for accurate measurement OR Fuse blown Rc Current stake C stake is w...

Страница 11: ...ions at rear Rc OK Resistance measurement settings 2P and 3P Selector switch Battery check setting voltage setting Test button Spike button Resistance scale Voltage scale Potential test spike resistan...

Страница 12: ...on Over under range indicator Output voltage indicator Ruptured fuse indicator Warning refer to user manual Battery charge indicator ART measurement indicator Possible Not possible Measuring range Cla...

Страница 13: ...tput voltage select button Test button Over under range indicator Output voltage indicator Ruptured fuse indicator Warning refer to user manual Battery charge indicator Measuring range Ground noise vo...

Страница 14: ...output voltage select button Test button Over under range indicator Output voltage indicator Ruptured fuse indicator Warning refer to user manual Battery charge indicator Measuring range Ground noise...

Страница 15: ...output voltage select button Test button Over under range indicator Output voltage indicator Ruptured fuse indicator Warning refer to user manual Battery charge indicator Measuring range Ground noise...

Страница 16: ...age indicator Ruptured fuse indicator Warning refer to user manual Battery charge indicator IART measurement indicator Possible Not possible Measuring range Clamp current presence indicator OK Not OK...

Страница 17: ...tage indicator Ruptured fuse indicator Warning refer to user manual Battery charge indicator ART measurement indicator Possible Not possible Measuring range Clamp current presence indicator OK Not OK...

Страница 18: ...ely after charging Nearly fully battery charge Some change remaining but instrument may auto power down at any time 0 battery charge remaining instrument will auto power down Note If the DET4TR2 and D...

Страница 19: ...electrical circuits 2 Move the slide on the terminal panel until the external AC DC adaptor socket is exposed 3 Remove the bung from the AC DC adaptor socket 4 Connect the AC DC adaptor and switch on...

Страница 20: ...the retaining screw Re enabling the battery charging circuit on the DET4TR2 and DET4TCR2 If the DET4TR2 and DET4TCR2 s rechargeable batteries are replaced with alkaline batteries the instrument may au...

Страница 21: ...itch instrument OFF and disconnect the instrument from any electrical circuits 2 The rear cover must not be opened if the test leads are connected 3 To remove the rear cover release the screw at the b...

Страница 22: ...c current of known magnitude into the system under test and measures the voltage developed across it as shown in Figure 3 The system resistance is a simple ratio as per Ohm s Law i e R V I Figure 3 sc...

Страница 23: ...m it is supposed to protect in the event of a power system fault The reason for this is that the injected test current will take all possible routes to ground and not all of it will necessarily flow t...

Страница 24: ...by the test voltage is I V Rx R1 R2 Rn It follows that as the resistance of the other electrodes in parallel approaches zero then the resistance measured approaches the value of the electrode under t...

Страница 25: ...asuring ground noise voltage 3 Set the rotary selector switch to the V position 4 The ground noise voltage reading will be displayed as shown in Figure 10 Figure 10 example ground noise voltage readin...

Страница 26: ...lector switch to the V position 4 The ground noise voltage reading will be indicated on the scaleplate as shown in Figure 12 Figure 12 example ground noise voltage reading Two terminal resistance meas...

Страница 27: ...terminal resistance reading will be displayed as shown in Figure 14 Figure 14 example two terminal resistance reading DET4 display shown Note The test voltage used to make the two terminal resistance...

Страница 28: ...will be inhibited if the ground noise voltage exceeds 40 V pk pk 7 The Rc LED will flash to indicate that the instrument is carrying out pre measurement checks 8 If the Rc LED does not illuminate fol...

Страница 29: ...hree terminal resistance 3 Set the rotary selector switch to the 3P position 4 Select the desired test voltage using the 25 V 50 V button 5 DET4TC2 and DET4TCR2 only Select the desired test frequency...

Страница 30: ...ure the rotary selector switch is in the OFF position 2 Select the desired test voltage using the procedure in the section on General Operating Instructions 3 Connect the instrument as shown in Figure...

Страница 31: ...ndicate that the instrument is carrying out pre measurement checks 11 If the Rc LED does not illuminate following the pre measurement checks this indicates that the C stake resistance exceeds the limi...

Страница 32: ...25 V 50 V button 5 DET4TC2 and DET4TCR2 only Select the desired test frequency using the Hz button 6 Press and release the TEST button by holding the TEST button the resistance measurement will be co...

Страница 33: ...the desired test frequency using the Hz button 6 Press and release the TEST button by holding the TEST button the resistance measurement will be continually updated 7 The instrument will perform pre...

Страница 34: ...round current 3 Close the MCC CLAMP around the conductor under test 4 Set the rotary selector switch to the A position 5 The ground current flowing in the conductor will be displayed as shown in Figur...

Страница 35: ...under test 4 Set the rotary selector switch to the 3P position 5 DET4TC2 and DET4TCR2 only Select the desired test frequency using the Hz button 6 Press and release the TEST button by holding the TES...

Страница 36: ...urements made Three terminal resistance measurement with lead null using ART DET4TC2 and DET4TCR2 In certain circumstances the resistance of the test leads connecting to the electrode under test may i...

Страница 37: ...strument will display the warning triangle above 2 A no ART testing is possible under this condition The instrument will display the warning triangle and an over range condition above 20 A no ART test...

Страница 38: ...could be lower than the actual resistance The resistance of the electrode or system must be verified by an alternative method Note The instrument will display the warning triangle above 2 A no Stakel...

Страница 39: ...instrument OFF using the selector switch 10 The instrument is now calibrated to the MCC CLAMP and the calibration stored in non volatile memory OPERATION OF THE TWO CLAMP CALIBRATION ADAPTOR 1 Ensure...

Страница 40: ...ALIBRATION ADAPTOR 1 Ensure the rotary selector switch is in the OFF position 2 Connect the instrument as shown 3 Set the rotary selector switch to the 2P 3P or 4P position 4 Press and release the TES...

Страница 41: ...mp 0 5 mA to 19 9 A Ground current accuracy 5 3 digits Ground voltage range 0 to 100 Vac Ground voltage accuracy 2 2 V Resistance range 0 01 to 2 k DET3TA DET3TC DET3TD 0 01 to 20 k DET4TD2 DET4TR2 0...

Страница 42: ...monitoring of protective measures Part 1 General requirements Part 5 Resistance to earth Dimensions 203 mm x 148 mm x 78 mm 8 in x 5 8 in x 3 in Weight 1 kg 2 2 lb MCC CLAMP SPECIFICATIONS Measuremen...

Страница 43: ...n Weight 0 5 kg approximately 1 2 lb CALIBRATION ADAPTOR ELECTRICAL SPECIFICATION Value 25 0 1 MECHANICAL SPECIFICATION Operating temperature 10 C to 50 C 14 F to 122 F 0 to 85 RH at 35 C 95 F Storage...

Страница 44: ...re for your Megger products Returning your product to Megger UK and USA service centres 1 When an instrument requires recalibration or in the event of a repair being necessary a Returns Authorisation...

Страница 45: ...ment Spent alkaline or NiMH AA batteries are classified as Portable Batteries and should be disposed of in the UK in accordance with Local Authority requirements For disposal of batteries in other par...

Страница 46: ...mited Unit 26 9 Hudson Avenue Castle Hill Sydney NSW 2125 Australia T 61 0 2 9659 2005 F 61 0 2 9659 2201 E ausales megger com Megger Limited 110 Milner Avenue Unit 1 Scarborough Ontario M1S 3R2 Canad...

Отзывы: