Fuse type
All instruments:
500 mA (F), HBC (50 kA, 600 V), 32mm x 6mm.
Megger part number: 25950-056.
Fuse replacement
Warning: Do not operate instrument with the battery cover removed.
Warning: Fitting an incorrect fuse type will reduce operator safety.
1.
To avoid the possibility of electric shock, switch instrument OFF and disconnect the instrument from any electrical
circuits.
2.
The rear cover must not be opened if the test leads are connected.
3.
To remove the rear cover, release the screw at the bottom of the cover and lift the cover upwards.
4.
To avoid the possibility of electric shock, do not press the TEST button when changing the fuse.
5.
Remove the ruptured fuse.
6.
Refit new fuse of the correct type.
7.
Replace the instrument back cover and secure by tightening the retaining screw