Megger CFL510F Скачать руководство пользователя страница 9

Operation

10

General Testing procedure

Ensure the correct test leads are firmly fitted into the
sockets of the instrument.

Switch on the instrument. The CFL510F will display
the start screen for a couple of seconds, followed by
a trace. The instrument will be set to the range and
velocity factor last used. If these settings are different
for the cable under test (C.U.T) then use the RANGE
and  VF keys to set the correct values. Refer also to
‘Velocity factor’ later in this text. 

Connection to Cable Under Test 

Connect the test lead to the cable under test.
Connection may be made to a live system with a
voltage to earth (ground) less than 300V with an
installation (over voltage) category of III or lower. This
means that the instrument may be connected to any
fixed wiring of a building installation, but not to
primary supply circuits such as overhead cables. 

A Fused Lead Set must be used to connect to
energised live systems. Refer to the accessories
section for options.

When selecting 25 cable impedance, an internal
50Hz/60Hz filter is automatically switched in.

Measuring distance to fault

Ensure that the whole length of the cable can be seen
on the display, and that the range selected is correct.

Find key

The Find key     searches for the furthest major
change of impedance. This may be a fault or the end
of the cable if this is a short or open circuit.

When found the event is displayed and the cursor
placed on the leading edge. The gain,  zoom and
cursor position may need to be adjusted for accurate
measurement. Note that Find does not adjust the
TxNull or Velocity Factor.

To enable ‘difficult to see’ faults to be identified, the
gain of the instrument can be adjusted. With the gain
at minimum small changes of cable impedance may
be overlooked. By increasing the gain the fault
becomes more obvious.

If no faults are obvious, adjust the gain until any major
reflections appear. Open and short circuits should
easily be identified. Partial faults can be less obvious.

If no significant reflections can be seen use the “TX
Null” to minimise the effect of the output pulse on any
potential “near end” faults (refer to section on TX
Null).

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание CFL510F

Страница 1: ...CABLE FAULT LOCATOR MEGGER CFL510F TDR1000 2 User Guide MEGGER www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 2: ...t are by necessity accessible The outer sheath for this connector is normally at SELV levels however under single fault conditions it may carry hazardous voltages The operator must therefore verify th...

Страница 3: ...to fault Menu TX Null Velocity Factor Pulse widths Techniques to improve accuracy Care and Maintenance 15 Specifications 16 Symbols used on the instrument are Caution Referto accompanying notes Equip...

Страница 4: ...ons are caused by changes in the cables characteristic impedance such as poor joints or discontinuities Faults showing an impedance higher than that of the cables normal impedance will cause a reflect...

Страница 5: ...nt to be read directly from the instrument Where the VF of a cable is not known but the length is the cursor can be set to the end of the cable and the VF on the TDR adjusted until the correct cable l...

Страница 6: ...leads loosen the battery cover retaining screws and remove the cover Lift out and disconnect the battery holder Replace the cells ensuring that correct polarity is observed shown on the battery holder...

Страница 7: ...ircuit of the TDR to that of the cable under test When correctly adjusted the majority of the displayed transmitted pulse can be nulled out allowing cable features close to the start of the cable to b...

Страница 8: ...value must match the VF of the cable under test to give an accurate distance measurement Menu Allows the user to change the units of measurement for Distance Velocity Factor Impedance and Test Rate Re...

Страница 9: ...internal 50Hz 60Hz filter is automatically switched in Measuring distance to fault Ensure that the whole length of the cable can be seen on the display and that the range selected is correct Find key...

Страница 10: ...beginning of the reflection The distance to the fault can then be directly read from the display NOTE The distance calculation is performed using the velocity factor set in the TDR If this velocity fa...

Страница 11: ...he LEFT or RIGHT cursor keys Available options Description Options Distance Unit The Units in which the distance to m metres or ft feet fault is displayed VF Format TDR velocity factor RATIO m us or f...

Страница 12: ...ight or as a distance per microsecond in ft us or m us If the exact length of cable is known and the reflection from the cable end is visible then an accurate velocity factor can be determined Locate...

Страница 13: ...and the most common and easily implemented methods Test the cable from both ends When fault finding a cable it is good practice to take measurements from both ends Particularly in the case of open cir...

Страница 14: ...e voltage between the terminals exceeds 300V Output pulse 5 volts peak to peak into open circuit Pulse widths determined by range and cable impedance 25 50 75 100 10m 7ns 7ns 7ns 7ns 30m 20ns 20ns 20n...

Страница 15: ...with Electromagnetic Compatibility Specifications Light industrial BS EN 61326 1 with a minimum performance of B for all immunity tests Mechanical The instrument is designed for use indoors or outdoo...

Страница 16: ...420 125 Fused prod and clip set 6111 218 Miniature Clip Test Lead Set 6231 652 User Guide 6172 659 For use on energised systems rated up to 300V CATIII use the fused prod and clip set AVO Part Number...

Страница 17: ...INTERNATIONAL Archcliffe Road PO Box 9007 4271 Bronze Way MEGGER SARL Dover Valley Forge Dallas 29 All e de Villemomble Kent CT17 9EN PA 19484 9007 TX 75237 1017 93340 Le Raincy England U S A U S A P...

Отзывы: