Operation
15
Except where otherwise stated, this specification
applies at an ambient temperature of 20C.
General
Ranges:
10m, 30m, 100m, 300m, 1000m, 3000m
(30ft, 100ft, 300ft, 1000ft, 3000ft, 10000ft)
Accuracy: 1% of range pixel at 0.67VF
[Note- The measurement accuracy is for the
indicated cursor position only and is conditional on
the velocity factor being correct.]
Resolution:
1% of range.
Input Protection:
This instrument complies with
IEC61010-1 for connection to
live systems up to 300V CAT
III. Fused leads must be used
if the voltage between the
terminals exceeds 300V.
Output pulse:
5 volts peak to peak into open
circuit. Pulse widths determined by range and cable
impedance:
25
Ω
50
Ω
75
Ω
100
Ω
10m
7ns
7ns
7ns
7ns
30m
20ns
20ns
20ns
20ns
100m
100ns
60ns
100ns
100ns
300m
300ns
120ns
170ns
300ns
1000m
1000ns
520ns
680ns
1000ns
3000m
3000ns
2020ns
2340ns
3000ns
Gain:
Set for each range with four user
selectable steps.
Velocity Factor: Variable from 0.01 to 0.99
in steps of 0.01
Impedance matching:
25, 50, 75, 100
Tx Null:
0 to 120
Refresh Rate:
Selectable from the menu.Once
per second or three times a second.
Power Down:
Automatic after 5 minutes with no
key press.