Megger CFL510F Скачать руководство пользователя страница 13

Operation

14

Pulse Widths

As the RANGE and IMPEDANCE matching of the
TDR are adjusted so the duration of the transmitted
pulse changes. The pulse widths range from 7ns to
3µs to overcome signal attenuation and enable the
instrument to see further down a length of cable. The
greater the range selected on the TDR, the wider the
transmitted pulse.

The accuracy of the “Distance to fault” is not affected
by the length of the pulse. However, if two or more
features exist close together (excluding open or short
circuits), then the second or subsequent feature may
be partially masked  by the reflection from the first
fault. Hence, for potential multiple features, the
instrument should be used with the shortest suitable
range, and so the smallest pulse width, that can see
both features.

For output pulse characteristics, refer to output pulse
data in the TDR Specification at the end of this guide. 

Techniques

To improve on the accuracy of the measurement,
numerous techniques can be used, depending on the
situation encountered. Not every situation can be
described, but the following points are effective and
the most common and easily implemented methods.

Test the cable from both ends

When fault finding a cable it is good practice to take
measurements from both ends. Particularly in the case
of open circuit faults, when the true end of the cable is
not visible. If the measurement is made from both
ends, then the combined answer should equal the
expected length of the cable. Even when the true end
of the cable is visible on the display, the reflections
after the fault may be too obscure to analyse clearly. In
this case, measurement from both ends yields a
clearer picture as well as improved accuracy.

It is also good practice to follow the cable route with
a cable tracer, as not all cable runs will be straight. It
can save a great deal of time if the exact route of the
cable is known as faults are often found at these
points and can be accredited to third party
intervention. 

Care and Maintenance

Other than replacing the batteries, the instrument has
no user serviceable parts. In case of failure it should be
returned to your supplier or an approved 
AVO INTERNATIONAL repair agent.

Cleaning the instrument should only be done by wiping
with a clean cloth dampened with soapy water or
Isopropyl Alcohol (IPA).

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание CFL510F

Страница 1: ...CABLE FAULT LOCATOR MEGGER CFL510F TDR1000 2 User Guide MEGGER www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 2: ...t are by necessity accessible The outer sheath for this connector is normally at SELV levels however under single fault conditions it may carry hazardous voltages The operator must therefore verify th...

Страница 3: ...to fault Menu TX Null Velocity Factor Pulse widths Techniques to improve accuracy Care and Maintenance 15 Specifications 16 Symbols used on the instrument are Caution Referto accompanying notes Equip...

Страница 4: ...ons are caused by changes in the cables characteristic impedance such as poor joints or discontinuities Faults showing an impedance higher than that of the cables normal impedance will cause a reflect...

Страница 5: ...nt to be read directly from the instrument Where the VF of a cable is not known but the length is the cursor can be set to the end of the cable and the VF on the TDR adjusted until the correct cable l...

Страница 6: ...leads loosen the battery cover retaining screws and remove the cover Lift out and disconnect the battery holder Replace the cells ensuring that correct polarity is observed shown on the battery holder...

Страница 7: ...ircuit of the TDR to that of the cable under test When correctly adjusted the majority of the displayed transmitted pulse can be nulled out allowing cable features close to the start of the cable to b...

Страница 8: ...value must match the VF of the cable under test to give an accurate distance measurement Menu Allows the user to change the units of measurement for Distance Velocity Factor Impedance and Test Rate Re...

Страница 9: ...internal 50Hz 60Hz filter is automatically switched in Measuring distance to fault Ensure that the whole length of the cable can be seen on the display and that the range selected is correct Find key...

Страница 10: ...beginning of the reflection The distance to the fault can then be directly read from the display NOTE The distance calculation is performed using the velocity factor set in the TDR If this velocity fa...

Страница 11: ...he LEFT or RIGHT cursor keys Available options Description Options Distance Unit The Units in which the distance to m metres or ft feet fault is displayed VF Format TDR velocity factor RATIO m us or f...

Страница 12: ...ight or as a distance per microsecond in ft us or m us If the exact length of cable is known and the reflection from the cable end is visible then an accurate velocity factor can be determined Locate...

Страница 13: ...and the most common and easily implemented methods Test the cable from both ends When fault finding a cable it is good practice to take measurements from both ends Particularly in the case of open cir...

Страница 14: ...e voltage between the terminals exceeds 300V Output pulse 5 volts peak to peak into open circuit Pulse widths determined by range and cable impedance 25 50 75 100 10m 7ns 7ns 7ns 7ns 30m 20ns 20ns 20n...

Страница 15: ...with Electromagnetic Compatibility Specifications Light industrial BS EN 61326 1 with a minimum performance of B for all immunity tests Mechanical The instrument is designed for use indoors or outdoo...

Страница 16: ...420 125 Fused prod and clip set 6111 218 Miniature Clip Test Lead Set 6231 652 User Guide 6172 659 For use on energised systems rated up to 300V CATIII use the fused prod and clip set AVO Part Number...

Страница 17: ...INTERNATIONAL Archcliffe Road PO Box 9007 4271 Bronze Way MEGGER SARL Dover Valley Forge Dallas 29 All e de Villemomble Kent CT17 9EN PA 19484 9007 TX 75237 1017 93340 Le Raincy England U S A U S A P...

Отзывы: