6. Connect the VOM between TP5 (the top plate of Q7) and ground; use a range of at
least 200 V dc. Set the OUTPUT KILOVOLTS switch to 3 kV. The meter should
read approximately 170 V dc, then drop back towards zero every six seconds.
Return the OUTPUT KILOVOLTS switch to ZERO START. A failure at this
point is probably due to SCR Q7, resistor R19, or transformer T3; check for shorts
and open circuits. The problem might also be in the capacitor bank C7; check for
damaged components and replace as necessary.
7. Check the secondary of pulse transformer T3 for an open circuit or a short. Check
current transformer T4, and check the circuitry feeding the meter.