11
2. Disconnect all ionising appliances from the controller.
3.
Reconnect the supply and switch ON the unit.
4.
Using a high voltage probe (RS type 610 281) and meter (RS type 610
950) measure the voltage on each of the output sockets. The reading
should be at minimum power 4kV at Max power 8 kV
5.
If the controller is found to be functional, test each of the attached products
independently following the instructions detailed in the appropriate Operation
and Maintenance manual.
If the 977v3 controller is found to be faulty please contact your local distributor or
Meech Static Eliminators Ltd.