Shelf Life of Reservoir Patch
: 2 years
Sterilization Method of Reservoir Patch
: By EO gas
Reservoir Volume
: 200 U (2 mL) (1 U=10 μL)
Insulin Type Used
: U-100
Basal Rate Range
:
0.00~ 10 U/h
(increment: 0.05 U/h)
Bolus Range
: 0.05 ~ 25 U (increment: 0.05 or 0.1 U)
Bolus Delivery Rate
: 0.05 U/2 s
Maximum Infusion Pressure and Occlusion Pressure Threshold:
15 psi
Maximum Time to Occlusion Alarm
:
Basal Delivery
(0.1 U/h): < 30 h
Basal Delivery
(1 U/h): < 3 h
Bolus Delivery
(3 U at 1.5 U/min): < 120 s
Bolus Volume after Occlusion Release
: < 3 U
Delivery Accuracy
:
Basal
: +/- 5% (at rates: 0.1~ 10 U/h)
Bolus
: +/- 5% (for all set values: 0.05 ~ 25 U)
Accuracy Test Results
(test cycle: 29 H, delivery rate: 1.0 U/H, average error:
0.40%):