1.4.5 Water
Both your Patch Pump and Sensor (including the installed Transmitter) are
waterproof to a depth of 2.5 meters (8 feet) for up to 60 minutes (IPX8). After
exposure to water, rinse the devices with clean water and dry them with a
towel.
Warning:
Do NOT expose your Patch Pump or Sensor (including the installed
Transmitter) to water at depths greater than 2.5 meters (8 feet) or for more
than 60 minutes. Check often to make sure the devices are securely attached
and in place.
Warning:
The PDM is not waterproof. Do NOT spill fluids on it or drop it into
fluids.
Warning:
The Patch Pump may not be able to deliver normally in water. The
Transmitter may not be able to send data normally in water.
Note:
Hot water may decrease Sensor life.
1.4.6 Storage
Store the Pump Base and Reservoir Patch at temperatures between -10°C (14°F)
and 55°C (131°F), and at humidity levels between 20% and 90% relative
humidity. Do NOT store the Pump Base and Reservoir Patch in direct sunlight,
extreme temperatures, or in very humid areas.
Store the Sensor at temperatures between 2°C (36°F) and 30°C (86°F), and at
humidity levels between 20% and 90% relative humidity for the length of the
Sensor’s shelf life. For temperatures greater than 30°C (86°F), the Sensor will
require cooled storage at temperatures no lower than 2°C (36°F). You may
Содержание A7 TouchCare
Страница 1: ...A7 TouchCare Insulin Management System User Guide ...
Страница 20: ...EXCLUDED INCLUDING THE WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ...
Страница 24: ... Glucose Sensor MD JY 006 or JY 016 Consumable Transmitter MD TY 012 USB charging cable ...
Страница 31: ...If you choose Skip the Bolus Calculation function will stay turned off ...
Страница 33: ...Maximum infusion with a single fault condition is 0 05U ...
Страница 48: ...UG882111GB 348300 Version 2 8 ...