Medtronic MIDAS REX LEGEND GOLD Скачать руководство пользователя страница 26

24

3.  (Foot Control) Pressure Relief Button Check

• 

Connect the foot control to a compressed gas source.

• 

Pressurize the foot control.

• Turn 

off  the compressed gas source.

• 

Press the Pressure Relief Button.

• Confi rm that the compressed gas has been released and the gage on the foot control drops to zero.

4.  (Foot Control) Finger/Foot Control Switch Functionality Check

• 

Place the switch in the 

 position.

• Confi rm the foot pedal is depressed.
• 

Place the switch in the 

 position.

• Confi rm the foot pedal is released.

5.  (Foot Control) Visually inspect the Legend and Triton port plugs for any signs of wear or damage and confi rm that they  

remain connected to the foot control.

6.  (Foot Control) Visually inspect the foot control pedal for any build-up of debris or indication of damage. Ensure the pedal 

moves freely.

7.  (Foot Control) Pedal Check

• 

Attach a Gold or Platinum motor to the foot control.

• 

Connect the foot control to a compressed gas source.

• 

Depress the foot control and run the motor for a short duration.

• 

Release the foot control and observe the motor. If the motor continues to run aft er the foot pedal control lever is 
released, 

DO NOT

 use and return the foot control to MPSS for service.

Содержание MIDAS REX LEGEND GOLD

Страница 1: ...MIDAS REX LEGEND Pneumatic High Speed System Preventive Maintenance Service Manual ...

Страница 2: ...d Maintenance 7 5th Scheduled Maintenance 8 6th Scheduled Maintenance 9 7th Scheduled Maintenance 10 8th Scheduled Maintenance 11 9th Scheduled Maintenance 12 10th Scheduled Maintenance 13 11th Scheduled Maintenance 14 12th Scheduled Maintenance 15 Section 2 Instructions for Completing Preventive Maintenance Steps 17 Scheduled Maintenance Intervals 1 2 3 5 6 7 9 10 11 18 Scheduled Maintenance Inte...

Страница 3: ... upon how often it is used and the conditions to which it is subjected such as cleaning and sterilization The demands on pneumatic systems can vary greatly from facility to facility Twelve maintenance intervals have been defined The time interval is determined by tool usage per year per system and has been identified as levels A B C or D in accordance with the following chart Maintenance Level No ...

Страница 4: ... in accordance with the Twelve Maintenance Intervals Your Midas Rex Legend Pneumatic System has a limited warranty for a period of 1 year from the date of purchase for Gold Gold Touch andPlatinumMotors Hoses PneumaticControlUnitwithPressureHose Regulator Adapters andInstrumentCasesand Trays The Attachments and Telescoping Tubes have a limited warranty for a period of 90 days from the date of purch...

Страница 5: ...ection 1 Preventive Maintenance Service Schedule 12 Maintenance Intervals Customer Performed Maintenance MPSS Performed Maintenance Date of System Purchase __________________ Maintenance Level ___________ ...

Страница 6: ...ping perforator and Jacobs Chuck drive attachment in this system NOTE Telescoping tubes are multi use disposable Discard if excessive heat or vibration is present Etching and color ring remain legible and distinguishable Attachment tube and base are clean and free of debris Attachment has no loose components Attachment tip shows no signs of wear or deformation Etching remains legible Connect the a...

Страница 7: ...oping perforator and Jacobs Chuck drive attachment in this system NOTE Telescoping tubes are multi use disposable Discard if excessive heat or vibration is present Etching and color ring remain legible and distinguishable Attachment tube and base are clean and free of debris Attachment has no loose components Attachment tip shows no signs of wear or deformation Etching remains legible Connect the ...

Страница 8: ...oping perforator and Jacobs Chuck drive attachment in this system NOTE Telescoping tubes are multi use disposable Discard if excessive heat or vibration is present Etching and color ring remain legible and distinguishable Attachment tube and base are clean and free of debris Attachment has no loose components Attachment tip shows no signs of wear or deformation Etching remains legible Connect the ...

Страница 9: ...____________________ Returned on ___________________ Contra Angled Attachment Replace Please call 817 788 6400 or toll free 1 800 433 7080 to order Ordered on ____________________ Air fittings and supply hose do not show any signs of blockage damage or wear Pressurize the foot control or regulator and confirm gage functionality NOTE The pressure gages should be placed on your facilities routine ca...

Страница 10: ...coping perforator and Jacobs Chuck drive attachment in this system NOTE Telescoping tubes are multi use disposable Discard if excessive heat or vibration is present Etching and color ring remain legible and distinguishable Attachment tube and base are clean and free of debris Attachment has no loose components Attachment tip shows no signs of wear or deformation Etching remains legible Connect the...

Страница 11: ...coping perforator and Jacobs Chuck drive attachment in this system NOTE Telescoping tubes are multi use disposable Discard if excessive heat or vibration is present Etching and color ring remain legible and distinguishable Attachment tube and base are clean and free of debris Attachment has no loose components Attachment tip shows no signs of wear or deformation Etching remains legible Connect the...

Страница 12: ...coping perforator and Jacobs Chuck drive attachment in this system NOTE Telescoping tubes are multi use disposable Discard if excessive heat or vibration is present Etching and color ring remains legible and distinguishable Attachment tube and base are clean and free of debris Attachment has no loose components Attachment tip shows no signs of wear or deformation Etching remains legible Connect th...

Страница 13: ...__________________ Returned on ___________________ NOTE Telescoping tubes are multi use disposable Discard if excessive heat or vibration is present Requires factory level service Please call 817 788 6440 or toll free 1 800 335 9557 to schedule Sent out on ____________________ Returned on ___________________ Contra Angled Attachment Replace Please call 817 788 6400 or toll free 1 800 433 7080 to o...

Страница 14: ...coping perforator and Jacobs Chuck drive attachment in this system NOTE Telescoping tubes are multi use disposable Discard if excessive heat or vibration is present Etching and color ring remain legible and distinguishable Attachment tube and base are clean and free of debris Attachment has no loose components Attachment tip shows no signs of wear or deformation Etching remains legible Connect the...

Страница 15: ...scoping perforator and Jacobs Chuck drive attachment in this system NOTE Telescoping tubes are multi use disposable Discard if excessive heat or vibration is present Etching and color ring remain legible and distinguishable Attachment tube and base are clean and free of debris Attachment has no loose components Attachment tip shows no signs of wear or deformation Etching remains legible Connect th...

Страница 16: ...scoping perforator and Jacobs Chuck drive attachment in this system NOTE Telescoping tubes are multi use disposable Discard if excessive heat or vibration is present Etching and color ring remain legible and distinguishable Attachment tube and base are clean and free of debris Attachment has no loose components Attachment tip shows no signs of wear or deformation Etching remains legible Connect th...

Страница 17: ... ____________________ Returned on ___________________ Contra Angled Attachment Replace Please call 817 788 6400 or toll free 1 800 433 7080 to order Ordered on ____________________ Air fittings and supply hose do not show any signs of blockage damage or wear Pressurize the foot control or regulator and confirm gage functionality NOTE The pressure gages should be placed on your facilities routine c...

Страница 18: ...16 ...

Страница 19: ...Section 2 Instructions for Completing Preventive Maintenance Steps 17 ...

Страница 20: ...ct the etching and ensure that it remains legible 6 Heat and Seal Check Attach and lock a straight or footed attachment onto the motor Check to ensure that there is a tactile feel and an audible click when the attachment is locked on the motor Connect the motor to a Foot Control unit attached to a compressed gas source Run the motor for 5 minutes Feel the motor case and the attachment base and ens...

Страница 21: ...n round with no evidence of deformation 5 Footed Visually inspect the foot The foot and or leg of the attachment should not be bent or damaged 6 Perforator and Jacobs Chuck Visually inspect the perforator collet for any signs of wear or deformation 7 Heat Check For telescoping attachment insert an appropriate telescoping tube into the base Attach and lock the attachment onto the motor Check to ens...

Страница 22: ...nect the motor to a Foot Control unit attached to a compressed gas source Run the motor for 5 minutes Feel the motor case the attachment base the knuckle angled area for angled attachment only and the attachment tip to ensure they are not uncomfortably hot to the touch 6 Contra Visually inspect the head to tube joint for the contra angled attachment for stability i e there is no loosening of this ...

Страница 23: ...libration date 3 Foot Control Pressure Relief Button Check Connect the foot control to a compressed gas source Pressurize the foot control Turn off the compressed gas source Press the Pressure Relief Button Confirm that the compressed gas has been released and the gage on the foot control drops to zero 4 Foot Control Finger Foot Control Switch Functionality Check Place the switch in the position C...

Страница 24: ...Check Attach a Gold or Platinum motor to the foot control Connect the foot control to a compressed gas source Depress the foot control and run the motor for a short duration Release the foot control and observe the motor NOTE If the motor continues to run after the foot pedal control lever is released DO NOT use and return the foot control to MPSS for service ...

Страница 25: ...tion Telescoping tubes are multi use disposable and do not receive scheduled maintenance Contra Angled Attachment The Contra angled attachment does not contain user or factory serviceable components and must be replaced at the 4th 8th and 12th maintenance intervals Please call 817 788 6400 or toll free 1 800 433 7080 to order Foot Control and Regulator 1 Visually inspect the air fittings and suppl...

Страница 26: ...is released 5 Foot Control Visually inspect the Legend and Triton port plugs for any signs of wear or damage and confirm that they remain connected to the foot control 6 Foot Control Visually inspect the foot control pedal for any build up of debris or indication of damage Ensure the pedal moves freely 7 Foot Control Pedal Check Attach a Gold or Platinum motor to the foot control Connect the foot ...

Страница 27: ...ts associated with any of the factory level service requirements are not included in the limited warranty Extended Maintenance Agreements are available Please refer to the first page insert for the Business Reply Card to obtain further information Telescoping tubes are multi use disposable and do not receive scheduled maintenance Contra Angled Attachment The Contra angled attachment does not conta...

Страница 28: ...26 Notes ...

Страница 29: ...gical Solutions Telephone 817 788 6400 Toll free 1 800 433 7080 Fax 817 788 6489 Repairs Medtronic Powered Surgical Solutions Telephone 817 788 6440 Toll free 1 800 335 9557 Fax 817 788 6441 EC Authorized Representative Medtronic B V Earl Bakkenstraat 10 6422 PJ Heerlen The Netherlands Patent Pending Pana is a trademark of NSK America Corporation 2014 Medtronic Inc All rights reserved Printed in U...