6. (Foot Control) Visually inspect the foot control pedal for any build-up of debris or indication of damage. Ensure the pedal
moves freely.
7. (Foot Control) Pedal Check
•
Attach a Gold or Platinum motor to the foot control.
•
Connect the foot control to a compressed gas source.
•
Depress the foot control and run the motor for a short duration.
•
Release the foot control and observe the motor.
NOTE:
If the motor continues to run aft er the foot pedal control lever is released,
DO NOT
use and return the foot
control to MPSS for service.